Functional description hi8014, Specifications hi8014 – Hanna Instruments HI 8014 User Manual

Page 3

Advertising
background image

4

5

FUNCTIONAL DESCRIPTION HI8014

ins rum n s

pH

pH meter

HI 8014

°C

mV/pH

AP

pH

°C

OFF

mV

2

1

3

4

1) BNC socket for combination pH electrodes.
2) Liquid Crystal Display (LCD).
3) Rotary Knob with the following selectable positions:

OFF

To switch the meter off (in any other position the
instrument is on).

°C

To display the manual temperature setting (through
°C trimmer).

pH

To display the pH value.

mV

To display the mV (ORP) readings when using an
ORP electrode or the mV equivalent to the pH values
when using a pH electrode.

4) Trimmers:

°C

To set temperature manually.

mV/pH For SLOPE calibration of pH.
AP

For OFFSET calibration of pH.

SPECIFICATIONS HI8014

HI 8014

Range

pH

0.00 to 14.00

mV

±1999

Resolution

pH

0.01

mV

1

Accuracy

pH

±0.01

(@20°C/68°F)

mV

±1

Typical EMC

pH

±0.05

Deviation

mV

±8

pH Calibration

Manual 2 point through
offset and slope trimmers

Offset Calibration

±1 pH

Slope Calibration

from 85 to 105%

Temperature

Manual from 0 to 100°C

Compensation

(32 to 212°F)

pH Electrode

HI 1230B combination, double-junction,

gel-filled, with BNC connector

and 1 m (3.3') cable (included)

Input Impedance

10

12

ohm

Battery Type

1 x 9V

Life

100 hours of continuous use

Environment

0 to 50°C (32 to 122°F);

max 95% RH non-condensing

Dimensions

185 x 82 x 53 mm

(7.3 x 3.2 x 2.1’’)

Weight

265 g (9.3 oz.)

Advertising