Hanna Instruments HI 9124 User Manual

Page 18

Advertising
background image

18

Typical Electrode Life

Ambient Temperature

1 – 3 years

90 °C

Less than 4 months

120 °C

Less than 1 month

Alkaline Error
High concentrations of sodium ions interfere with readings in alkaline
solutions. The pH at which the interference starts to be significant
depends upon the composition of the glass. This interference is called
alkaline error and causes the pH to be underestimated. Hanna’s glass
formulations have the indicated characteristics.

1.0 Mol L

-1

Na

+

0.1 Mol L

-1

Na

+

Sodium Ion Correction for Glass at 20

-25 °C

Concentration

pH

Error

13.00
13.50
14.00
12.50
13.00
13.50
14.00

0.10
0.14
0.20
0.10
0.18
0.29
0.40

The resistance of glass electrodes partially depends on temperature. The
lower the temperature, the higher the resistance. It takes more time for
the reading to stabilize if the resistance is higher. In addition, the response
time will suffer to a greater degree at temperatures below 25 °C.

Since the resistance of the pH electrode is in the range of 50 – 200
Mohms, the current across the membrane is in the pico Ampere range.
Large currents can disturb the calibration of the electrode for many
hours.
For these reasons high humidity environments, short circuits and static
discharges can be detrimental to a stable pH reading.
The pH electrode’s life also depends on the temperature. If constantly
used at high temperatures, the electrode life is drastically reduced.

TEMPERATURE CORRELATION

TEMPERATURE CORRELATION

TEMPERATURE CORRELATION

TEMPERATURE CORRELATION

TEMPERATURE CORRELATION

FOR

FOR

FOR

FOR

FOR ppppp H SENSITIVE GLASS

H SENSITIVE GLASS

H SENSITIVE GLASS

H SENSITIVE GLASS

H SENSITIVE GLASS

Advertising
This manual is related to the following products: