Hanna Instruments HI 83141V User Manual

Hanna Instruments Measuring instruments

Advertising
background image

Instruction Manual

HI 83141V

Complete Portable Kit

for pH, mV & Temperature

Measurements in

Wine, Must and Grape Juice

w w w . h a n n a i n s t . c o m

WARRANTY

All Hanna Instruments meters are warranted for two years against
defects in workmanship and materials when used for their intended pur-
pose and maintained according to instructions.
The electrodes and the probes are warranted for a period of six
months. This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed main-
tenance are not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of purchase,
serial number and the nature of the failure.
If the repair is not covered by the warranty, you will be notified of the
charges incurred. If the instrument is to be returned to Hanna Instruments,
first obtain a Returned Goods Authorization Number from the Customer
Service department and then send it with shipment costs prepaid.
When shipping any instrument, make sure it is properly packaged for
complete protection.

Dear Customer,
Thank you for choosing a Hanna product.
Please read carefully this instruction manual before using the
meter. If you need additional technical information, do not
hesitate to e-mail us at [email protected].
This instrument is in compliance with the

directives.

Remove the kit from the packing material and examine it
carefully to make sure that no damage has occurred during
shipment. If there is any damage, notify your dealer.
The kit includes:
• HI 83141 portable pH meter
• HI 1048B pH electrode
• HI 7669AW temperature probe
• pH 3 and pH 7 buffer solutions (20 mL sachet each)
• HI 700635 and HI 700636 specific cleaning solutions
• electrode refilling solution (30 mL bottle) and syringe
• 9V battery, screwdriver and instructions.
Note: Conserve all packing material until the instrument has

been observed to function correctly. Any defective item
must be returned in its original packing.

PRELIMINARY EXAMINATION

FUNCTIONAL DESCRIPTION

SPECIFICATIONS

1) BNC connector for pH (or ORP) electrode.
2) Liquid Crystal Display.
3) mV key, to display the mV (ORP) readings when using

an ORP electrode or the mV equivalent to pH values
when using a pH electrode.

4) Trimmer for pH OFFSET calibration.
5) Trimmer for pH SLOPE calibration.
6) °C key, to display the temperature measurement.
7) pH key, to display the pH value.
8) ON/OFF key.
9) Socket for HI 7669AW temperature probe.

Range

0.00 to 14.00 pH / ±1999 mV

0.0 to 100.0°C

Resolution

0.01 pH / 1 mV / 0.1°C

Accuracy

(@20°C/68°F)

±0.01 pH / ±1 mV / ±0.4°C

Typical EMC Deviation

±0.07 pH / ±5 mV / ±1°C

pH Calibration

Manual, 2 point, through

offset (±1 pH) and slope (85 to 105%) trimmers

ORP Calibration

Factory calibrated

Temperature Calibration Factory calibrated
Temperature Compensation

Automatic, from 0 to 70°C

Probes (included)

HI 1048B pH-electrode

HI 7669AW temperature probe

Battery Type

1 x 9V, alkaline

Battery Life

100 hours of continuous use

Environment

0 to 50°C (32 to 122°F);

RH max 95% non-condensing

Dimensions

164 x 76 x 45 mm (6.5 x 3.0 x 1.8")

Weight

180 g (6.3 oz.)

ACCESSORIES

HI 1048B

Glass body, double-junction, refillable pH elec-
trode with BNC connector and 1 m (3.3’) cable

HI 3131B

Glass-body, refillable, platinum ORP electrode
with BNC connector and 1 m (3.3’) cable

HI 7669AW Temperature probe with 1 m (3.3’) cable
HI 5003

pH 3.00 buffer solution, 500 mL bottle

HI 5068

pH 6.86 buffer solution, 500 mL bottle

HI 5007

pH 7.01 buffer solution, 500 mL bottle

HI 70300L

Storage solution, 500 mL bottle

HI 70635L

Cleaning solution for wine deposits, 500 mL

HI 70636L

Cleaning solution for wine stains, 500 mL

HI 7082

Electrolyte solution, 3.5M KCl, 4 x 30 mL bottle

HI 7091L

Reducing pretreatment solution, 500 mL

HI 7092L

Oxidizing pretreatment solution, 500 mL

HI 731326

Calibration screwdriver (20 pcs)

Hanna Instruments reserves the right to modify the design, construction

and appearance of its products without advance notice.

Recommendations for Users

Before using this product, make sure that it is entirely suitable for the environment in which
it is used. Operation of this instrument in residential area could cause unacceptable interference
to radio and TV equipment, requiring the operator to take all necessary steps to correct
interference.

The glass bulb at the end of the electrode is sensitive to electrostatic discharges. Avoid
touching this glass bulb at all times.

To maintain the EMC performance of equipment, the recommended cables noted in the
user's manual must be used. Any variation introduced by the user to the supplied equipment
may degrade the instruments' EMC performance.

To avoid electrical shock, do not use these instruments when voltages at the measurement surface
exceed 24 Vac or 60 Vdc.

To avoid damage or burns, do not perform any measurement in microwave ovens.

IST83141V 07/05