Hanna Instruments HI 83140 User Manual

Meter with probe check

Advertising
background image

Instruction Manual

HI 83140

Portable pH/mV/ºC

meter

with Probe Check

Hanna Instruments reserves the right to modify the design, con-

struction and appearance of its products without advance notice.

WARRANTY

All Hanna Instruments meters are warranted for two years against
defects in workmanship and materials when used for their intended
purpose and maintained according to instructions. The electrodes and
the probes are warranted for a period of six months. This
warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of pur-
chase, serial number and the nature of the failure. If the repair is not
covered by the warranty, you will be notified of the charges incurred. If
the instrument is to be returned to Hanna Instruments, first obtain a
Returned Goods Authorization Number from the Customer Service de-
partment and then send it with shipment costs prepaid. When shipping
any instrument, make sure it is properly packaged for complete protec-
tion.

All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner.

Dear Customer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for correct opera-
tion of the meter. Please read it carefully before using the
meter. If you need additional technical information, do not
hesitate to e-mail us at [email protected].
This instrument is in compliance with directives.

Remove the instrument from the packing material and
examine it carefully to make sure that no damage has
occurred during shipment. If there is any damage, notify
your Dealer.
Each meter is supplied with:

• HI 1217-6, amplified, combination, double-junction,

plastic-body, gel pH electrode with built-in tempera-
ture sensor;

• HI 70007 pH 7.01 sachet (1 pc)
• HI 70004 pH 4.01 sachet (1 pc)
• HI 700661 Cleaning Solution sachet (2 pcs)
• HI 50021 Check Solution sachet (2 pcs)
• Instruction Manual
• Calibration screwdriver
• 1 x 9V alkaline battery.

Note: Conserve all packing material until the instrument

has been observed to function correctly. Any defective
item must be returned in its original packing.

PRELIMINARY EXAMINATION

GENERAL DESCRIPTION

HI 83140 is an hand-held pH/mV/°C meter designed for
simplicity of use in taking pH, mV (ORP) and temperature
measurements.
The pH, mV and °C ranges are easily selected using a
membrane keyboard on the front panel.
Calibration procedure is very simple and adjustments are
easily made with two trimmers on the front panel.
The meter is also provided with Probe Check feature, which
allows the user to easily check the probe calibration status
at any time.
The built-in temperature sensor in the pH electrode allows
automatic temperature compensation of pH readings.
Moreover HI 83140 has been designed with a low battery
detector feature and a lightweight ABS plastic housing.

FUNCTIONAL DESCRIPTION

SPECIFICATIONS

1) DIN connector for combination pH or ORP electrode.
2) Liquid Crystal Display.
3) Green and Red LED’s for Probe Check.

4) Check key.
5) mV key, to display the mV (ORP) readings when

using an ORP electrode or the mV equivalent to pH
values when using a pH electrode.

6) OFFSET trimmer for pH calibration.
7) SLOPE trimmer for pH calibration.
8) °C key, to display the temperature measurement.
9) pH key, to display the pH value.

10) ON/OFF key.

Range

pH

0.00 to 14.00

mV

±1999

°C

0.0 to 100.0

Resolution

pH

0.01

mV

1

°C

0.1

Accuracy

pH

±0.01

(@20°C/68°F)

mV

±1

°C

±0.4

Typical EMC

pH

±0.05

Deviation

mV

±5

°C

±0.5

pH Calibration

Manual, 2-point through trimmers

Temperature

Automatic from 0 to 70°C (32 to 158°F)

Compensation
Electrode (*)

HI 1217-6 pH-electrode with built-in

(included)

temperature sensor and DIN connector

Battery Type

1 x 9V (IEC 6LR61) alkaline

Battery Life

approx. 150 hours of continuous use

Environment

0 to 50°C (32 to 122°F);
max 95% RH non-condensing

Dimensions

164 x 76 x 45 mm (6.5 x 3.0 x 1.8")

Weight

250 g (8.8 oz.)

(*) The max. operating temperature of the HI 1217-6 pH electrode is 80ºC.
To perform measurements at higher temperature use adequate electrodes.

BATTERY REPLACEMENT

When the battery becomes weak, the
meter displays a blinking additional deci-
mal point on the left side of the LCD.
When the low battery indicator appears only a few hours of
battery life is remaining. A low battery level may also result
in unreliable measurements. It is recommended to replace
the battery soon.
Unscrew the 3 screws on the back of the meter, remove the
battery cover and replace the battery while paying atten-
tion to its polarity.
Replacement must only take place in a non-hazardous area
using an alkaline 9V battery.