Hanna Instruments HI 4421 User Manual

Page 38

Advertising
background image

38

Make sure the instrument has been calibrated before taking DO measurements.

DIRECT MEASUREMENT
To measure the DO of a sample using the Direct reading mode:
• Press

and then

to select DO measure mode.

• Select the

Direct

reading mode (see

DO

Setup

).

• Submerge the DO probe and tap

it repeatedly to remove any

air bubbles that may be trapped inside the sleeve. Allow time

for the reading to stabilize.

• The measured DO value will be displayed together with the

temperature and pressure values.

Notes: • For accurate DO measurements, a water movement of

0.3 m/s is required. This is to ensure that the oxygen-

depleted membrane surface is constantly replenished.

A moving stream will provide adequate circulation.

• If the reading is out of range, “-----” will be displayed.

DIRECT/AUTOHOLD MEASUREMENT
To measure DO of a sample using the Direct/AutoHold reading mode:
• Select the Direct/AutoHold reading mode (see

DO

Setup

).

• If pressing

, the “AutoHold” indicator will start blink-

ing on the display until the stability criterion is reached. The DO
value will be frozen on the display, along with “AutoHold”
indicator.

• To return to normal measure mode press

.

DO MEASUREMENT

DO MEASUREMENT

DO MEASUREMENT

DO MEASUREMENT

DO MEASUREMENT