Alarm
This option allows the user to define two alarm limits for the measurements.
Alarm State
The following options are available:

Disabled

– the alarm will be disabled.

Inside Limits

– the alarm will notify the user when the measured value is inside the preset limits.

Outside Limits

– the alarm will notify the user when the measured value is outside the preset limits.

To set the alarm state:
• Use

or

to highlight the

Alarm State

option.

• Press

and use

or

to highlight the

desired option.

• Press

to confirm your selection or press

to

cancel operation.

Alarm Limits

This option allows the user to set the alarm limits for the measured

value.

Note: The alarm high value cannot be lower than the alarm low

value.

To set the alarm limits:
• Highlight the

Alarm Limits

option and then press

.

• Use

or

to select the low / high alarm limit and

then press

.

• Use

or

to increase / decrease the selected

alarm value.

• Press

to return to the

Alarm

options.