• When in MTC mode, if pressing

after entering

calibration and while a custom pH calibration buffer is
selected, a pop-up menu will be displayed on the LCD
in which the custom buffer and the temperature value
can be ajusted by pressing

and then

or

keys. Press

to save the modified

value and then

/

to select next/

previous value to be adjusted.

• When in ATC mode, if pressing

after entering

calibration and while a standard pH buffer (with a
x.xxx resolution) or a custom pH buffer is selected, a
pop-up menu will be displayed on the LCD in which
the buffer value can be ajusted using

or

. Press

to save the new buffer value.

• If the Automatic buffer entry type has been selected for the calibration procedure, the instrument

will automatically select the closest buffer to the measured pH value from the edit buffer group
(see pH Setup for details, page 20).

• If the Semiautomatic buffer entry type has been selected for the calibration procedure, the instrument

will automatically select the closest buffers to the measured pH value from all the available buffers
and the used buffer value can be selected with

or

.

CALIBRATION MESSAGES
• Wrong buffer. Please check the buffer: this message appears when the difference between the pH

reading and the value of the selected calibration buffer is significant. If this message is displayed, check
if you have selected the appropriate calibration buffer.

• Wrong buffer temperature: this message appears if the buffer temperature is out of the defined buffer

temperature range.

• Clean the electrode or check the buffer. Press <Accept> to update calibration: this message alerts the

user that some dirt or deposits could be on the electrode. Refer to the electrode Cleaning Procedure (see page 65).

• Slope too low. Please check the buffer / Slope too high. Please check the buffer: these messages appear

if the current slope is under 80% or over 110% of default slope. Recalibrate the instrument using fresh buffers.

• Slope too low. Press <Clear Cal> to clear old calibration / Slope too high. Press <Clear Cal>

to clear old calibration: these messages appear as a result of an erroneous slope condition. Follow
displayed instructions.