Hanna Instruments HI 4008 User Manual

Page 12

Advertising
background image

12

Procedure
1) Follow sections VIII and IX to prepare sensors for

measurement.

2) Follow section VI to prepare standards / solution.

Standards should bracket and fall within the range of
interest.
Two mL HI 4000-00 ISA is added to 100 mL of both
samples and standards. Add stir bar and mix before
taking measurements.

3) Follow section VII; General Guidelines to optimize test

set-up.

4) During calibration it is best to start with lower concen-

tration samples first. Wait for a stable measurement
before recording values. Slightly longer equilibra-
tions are required at lower concentrations .

5) To prevent carry over and contamination of samples,

rinse sensors with deionized and dab dry with
absorbant laboratory tissue between samples.

XII.

XII.

XII.

XII.

XII. Other Measurement T

Other Measurement T

Other Measurement T

Other Measurement T

Other Measurement Techniques

echniques

echniques

echniques

echniques

Known Addition (for Cu

2+

)

An unknown concentration can be determined by adding a
known amount (volume and concentration) of measured
ion to a known volume of the sample. A mV value is taken
before and after the addition of standard, and using the
equation provided, the unknown concentration is found.

Typical Linearity for HI 4008 and HI 4108 Cupric

Electrodes

0

20

40

60

80

100

120

140

160

180

0.5

1

1.5

2

2.5

3

3.5

4

4.5

5

5.5

6

6.5

7

-Log[M]

mV