Hanna Instruments HI 250 User Manual

Page 17

Advertising
background image

17

LOG

DATE

LOG

mV

• The date on the primary LCD, along with “DATE” tag.

• The calibration offset on the primary LCD and “OFS” message on the

secondary LCD.

• The calibration slope on the primary LCD and “SLP” message on the

secondary LCD.

Note: • If in Relative mV RECALL mode, the instruments will display “----”

message on the primary LCD (HI 251 only).

• When the time, date, year, offset or slope is displayed, if SET/CLR

is pressed and hold down, the secondary LCD will display the record

number.

• The “dEL” message on the primary LCD and the selected record on the

secondary LCD, along with ”LOG” tag.

Note: • If one of the ARROW keys is pressed while “dEL” and record

number is displayed, the next/previous record number is selected.

• If LOG is pressed, the secondary LCD will display “ALL” message.

• Pressing LOG again, the secondary LCD will display the record

number.

• Press SET/CLR to delete the selected or all records.

• If “del ALL” option was selected, all records are deleted and the

instrument returns to measurement mode.

Note: After LOG is pressed while in measurement mode or a selected record is

deleted, the instrument will display the amount of free log space for about

one second (e.g. 25 records free).

Press RCL to leave RECALL mode at any time.