Hanna Instruments HI 2216 User Manual

Page 29

Advertising
background image

29

LOG INTERVAL
Press

CAL when log interval is displayed. The log interval and “CFM” will

start blinking (“

MAn” for log on demand, “StAb” log on stability,

interval in seconds or minutes).
Press the

ARROW keys to change the custom buffer value.

Press

CFM to confirm the selection.

Press

CAL to escape without saving.

EXPIRED CALIBRATION ALARM
Press

CAL when the calibration time-out is displayed. Calibration time-out

(“

OFF” or “1” to “7” days) and “CFM” will start blinking.

Press the

ARROW keys to change the calibration time-out value.

Press

CFM to save the modified calibration time-out value.

Press

CAL to escape without saving.

FIRST CUSTOM BUFFER

Press

CAL when “cb1” is displayed. The custom buffer (disabled – “no”

or “

0” to “16” pH) and “CFM” will start blinking.

Press the

ARROW keys to change the custom buffer value.

Press

CFM to save the modified custom buffer value.

Press

CAL to escape without saving.

Note: To delete custom buffer press CAL then press CLR key.

SECOND CUSTOM BUFFER

Press

CAL when “cb2” is displayed. The custom buffer (disabled – “no”

or “

0” to ”16” pH) and “CFM” will start blinking.

Press the

ARROW keys to change the custom buffer value.

Press

CFM to save the modified custom buffer value.

Press

CAL to escape without saving.