SET

Is equivalent to pressing SET

PRT

Is equivalent to pressing PRINT (HI 123 only)

CLR

Is equivalent to pressing CLR

OFF

Is equivalent to pressing OFF

AHD

Is equivalent to pressing AutoHOLD

CHR xx

Change the instrument range according with the parameter
value (xx):

• xx=00 pH range/0.001 resolution, pH channel
• xx=01 pH range/0.01 resolution, pH channel
• xx=03 mV range, pH channel
• xx=04 Relative mV range, pH channel
• xx=13 mV range, ISE channel
• xx=14 Relative mV range, ISE channel
• xx=15 ISE range, ISE channel

The meter will answer for these commands with:

<STX> <answer> <ETX>

where:

<STX> is 02 ASCII code character (start of text)
<ETX> is 03 ASCII code character (end of text)
<answer>:

<ACK> is 06 ASCII code character (recognized command)
<NAK> is 21 ASCII code character (unrecognized command)
<CAN> is 24 ASCII code character (corrupted command)

COMMANDS REQUIRING AN ANSWER

RAS

Causes the instrument to send a complete set of readings
in according with the current range:

• pH, mV and temperature reading on pH range.
• mV and temperature reading on mV range.
• Rel mV, absolute mV and temperature reading on

Rel mV range.

Note: The temperature is sent with two decimals and in Celsius degrees,

even if the set temperature unit was Fahrenheit degrees.

DA?

Requests the date.

TI?

Requests the time.

MDR

Requests the instrument model name and firmware code.

GLP

Requests the calibration data record.

PAR

Requests the setup parameters setting.

NSL

Requests the number of logged samples.

?ML

Requests the information about AutoLOG.