Hanna Instruments HI 981401N User Manual

Page 2

Advertising
background image

Always replace the protective cap after use.

DO NOT USE DISTILLED OR DEIONIZED WATER FOR
STORAGE PURPOSES.

• If the electrode has been left dry, soak the tip in a

storage (HI 70300) or pH7 (HI 7007) solution for at
least one hour to reactivate it.

• To minimize clogging and provide longer life for the pH

electrode, it is recommended to clean it monthly.

Immerse the tip of the electrode in HI 7061 for one
hour, then rinse it with tap water.

TAKING pH MEASUREMENTS

• Turn the meter on by connecting the 12 Vdc power

adapter to the meter and to the mains.

• In addition, the meter is supplied

with a differential input and
HI 1283 stainless steel probe to
prevent potential grounding prob-
lems. Remove the protective cap
from the pH electrode and immerse
the tips (4 cm/1½”) of both pH
electrode and grounding probe into
the sample.

• The LCD will show the pH value. Allow

the reading to stabilize and the meter
will start continuous monitoring.

Note: To prolong the electrode life, after switching the meter

off, remove the electrode from the solution.

For the greatest accuracy, frequent calibration of the instru-
ment is recommended. In addition, the instrument must be
recalibrated whenever:

a) The pH electrode is replaced
b) After testing aggressive chemicals
c) Extreme accuracy is required
d) At least once a month

PREPARATION

Pour small quantities of pH 7.01 (HI 7007) and pH 4.01
(HI 7004) solution into two clean beakers.
For accurate calibration use two beakers for each buffer
solution, the first one for rinsing the tip of the electrode and
the second one for calibration.

As second point for calibration, pH 4.01 buffer (HI 7004) is
recommended for measuring acidic samples, while pH 10.01
(HI 7010) should be used if the samples to be tested are
alkaline.

CALIBRATION PROCEDURE

• Remove the protective cap from the

electrode, rinse and immerse it in
pH 7.01 buffer solution together
with the grounding probe. Stir gen-
tly and wait a couple of minutes for
the reading to stabilize.

Note: the electrode should be sub-

merged approximately 4 cm
(1½”) in the solution.

CALIBRATION

• With the supplied screwdriver,

adjust the “pH 7” calibration
trimmer until the display shows
“7.0” pH.

• Rinse and immerse pH electrode and grounding probe in

pH 4.01 (or pH 10.01) buffer solution.

• Stir gently, wait a couple of

minutes and adjust the “pH
4” trimmer until the LCD
shows “4.0” pH (or “10.0”
pH).

The calibration is now complete.

4 cm

(1½")

CE DECLARATION OF CONFORMITY

Recommendations for Users
Before using this product, make sure that it is entirely suitable for the environment in
which it is used. Operation of this instrument in residential areas could cause unaccept-
able interferences to radio and TV equipment.
The glass bulb at the end of the electrode is sensitive to electrostatic discharges. Avoid touching
this glass bulb at all times. During operation, ESD wrist straps should be worn to avoid possible
damage to the electrode by electrostatic discharges.
Any variation introduced by the user to the supplied equipment may degrade the instrument’s
EMC performance. To avoid electrical shock, do not use this instrument when voltages at the
measurement surface exceed 24 Vac or 60 Vdc.
To avoid damages or burns, do not perform any measurement in microwave ovens.

ACCESSORIES

HI 1283 *

Stainless steel grounding probe with 2 m
(6.6’) cable

HI 1286

Double junction, plastic body pH electrode
with 2 m (6.6’) cable and BNC connector

HI 70004P

pH 4.01 solution, 20 mL sachet (25 pcs)

HI 70007P

pH 7.01 solution, 20 mL sachet (25 pcs)

HI 70010P

pH 10.01 solution, 20 mL sachet (25 pcs)

HI 7004L

pH 4.01 solution, 500 mL bottle

HI 7007L

pH 7.01 solution, 500 mL bottle

HI 7010L

pH 10.01 solution, 500 mL bottle

HI 70300L

Electrode storage solution, 500 mL bottle

HI 7061L

Electrode cleaning solution, 500 mL bottle

HI 710005

12 Vdc power adapter, US plug

HI 710006

12 Vdc power adapter, European plug

* To be replaced by authorized technical personnel only.