Hanna Instruments HI 98713 User Manual

Page 15

Advertising
background image

28

29

Press GLP to enter or exit GLP data consulting.
Several functions are available when in GLP menu.

Press READ to scroll the following GLP data:

• The last calibration date, in YYYY.MM.DD

format. If no calibration was performed, the
factory calibration message, “F.CAL”, will be
displayed on the LCD.

• The time of the last calibration in hh:mm

format.

• First calibration point: 0.00 FNU if skipped or

the actual read value (e.g. 0.01 FNU).

• Second calibration point.

GOOD LABORATORY PRACTICE (GLP)

GOOD LABORATORY PRACTICE (GLP)

GOOD LABORATORY PRACTICE (GLP)

GOOD LABORATORY PRACTICE (GLP)

GOOD LABORATORY PRACTICE (GLP)

• Measurement date in YYYY.MM.DD format.

• Measurement time in hh:mm format.

• Delete the last record panel (only for last re-

cord).

• Delete all records.

DELETE LAST RECORD
To log other values, the last record or all records
have to be deleted.
• To delete the last record, press SETUP/DEL while

in delete last records panel.

• The instrument asks for confirmation and if

LOG/CFM is pressed, the last record is deleted.
To abort the delete function, press READ
instead of LOG/CFM.

• After the record is deleted, the instrument goes

immediately to the first panel of the previous
record. If the log becomes empty, dashes will
be displayed for one second and the instrument
will return to idle mode.

DELETE ALL RECORDS
To delete all records, scroll the log until delete all
records panel is displayed on the LCD.

• To delete all records press SETUP/DEL while in

delete all records panel.

• The instrument asks for confirmation and if

LOG/CFM is pressed, all records are deleted.
To abort the delete function, press READ
instead of LOG/CFM.

• After all records are deleted, dashes are

displayed for one second and the instrument
returns to idle mode.

The GLP feature allows the user to view last calibration data. Also, the user calibration can be
deleted.

or

or