Hanna Instruments HI 98703 User Manual

Page 9

Advertising
background image

9

8)

ON/OFF, press to turn the instrument ON/OFF. If no key is pressed for more than
15 minutes, the instrument automatically shuts off.

9)

GLP

, press to enter/exit GLP feature. In SETUP it is used to increase the set values. In

Log Recall it is used to select a newer record (scroll up).

10)

AVG

, press to set the average reading mode ON/OFF. In SETUP it is used to decrease the

set values. In Log Recall it is used to select an older record (scroll down).

11)

CAL, press to enter/exit calibration. During setup it is used to start/stop editing a
parameter.

12)

LIGHT, press to turn ON/OFF the backlight.

13)

RCL, press to enter/exit viewing log content.

14)

LOG/CFM, press to save the log records or to confirm the selected option.

15)

READ , press to start a measurement. Press and hold, to make a continuous measurement.
In Log Recall it is used to see the content of a record. In GLP it is used to see all available
informations. In SETUP, during date or time editing, it is used to move the focus on the next
setting item.

16)

SETUP/DEL, press to enter/exit setup. The DEL function is available in Log Recall to delete
one or all records. In GLP it is used to delete the user calibration.

CONNECTORS DESCRIPTION

17) AC adapter connector, used to connect an external AC Adapter.
18) RS232 connector, used to transfer data through the RS232 connection. Use

HI 920011

serial cable to connect to the PC.

19) Tag reader connector. Touch the tag with the connector to read the location identification

number during logging.

20) USB connector, used to transfer data to the PC.

Advertising