• If the calculated calibration coefficients are

outside a certain range the “

CAL Err” message

is displayed.

CALIBRATION DELETION
HI 93414 is factory calibrated. It is possible to
restore factory calibration by deleting the last
performed calibration.
To delete last calibration, follow the next steps:

• Enter the GLP feature by pressing

RCL/GLP

for three seconds. The date of the last
calibration will be displayed on the LCD.

• Press

READ/TIMER to see the information

related to calibration. The last panel is the one
with delete calibration.

• Press

SETUP/DEL to delete the current

calibration. The instrument will display “

dEL

donE” for a second and the calibration is
deleted, then the instrument will automatically
return to measurement mode.