Setup mode – Hanna Instruments HI 98701 User Manual

Page 5

Advertising
background image

9

8

PASSWORD PROTECTION

Setting of the GLP parameters (calibration alarm time-out, instrument
ID code, time and date) can be password protected. If password is set
to a value different from 0000 (factory setting), the user will be asked
to enter the password.
• Select the desired GLP parameter code.
• Enter the password by the arrow keys.

• Press the ALT and CFM keys to

confirm.

• If password is wrong the meter will return to the function selection

mode without any warning message.

• If password is correct, the meter provides access to the parameter.

PARAMETERS SETTING

• Once the parameter code has been entered, the appropriate mes-

sage will scroll across the LCD for a few seconds.

• The current value of the selected parameter on the upper LCD and

the parameter code on the lower LCD will be displayed. The first
digit will blink if the parameter can assume continuous values. All
the digits will blink if the parameter can assume only a fixed set of
values.

• Enter the new value using the arrow keys.
• Press ALT and CFM to confirm the value.
The following table lists the setup codes along with the description of
the specific setup items, their valid values and the factory settings
(default):

Code

Valid values

Default

00 Lot data printing/scrolling

1

00÷16

00

01 Print lots data summary

1

02 Printer enable

1

On(enabled); Off(disabled)

On

03 Logging interval

1,2,5,10,15,30,60,120,180

1

05 Log on demand delete

1

or Reset sample number

2

06 Timed data delete

1

Setup can be used to view data regarding instrument status (e.g.
battery charge) or GLP data (e.g. calibration date) or to view or print
the logged data. It also allows the user to change the meter param-
eters (e.g. time) and to gain access to stored data.

• To enter this mode ensure the meter is

not logging and then press the ALT and
FNC keys.

• The scrolling message "Insert the function code or press "ALT" -

"FNC" to escape" in the upper LCD and the indication "F 00" with
the first digit blinking in the lower LCD will be displayed.

• Enter the first digit of the code of the parameter you want to set

using the or key and pass to the next digit with

. The

second digit will start blinking.

• Enter the second digit using the or

key.

• Press ALT and CFM to confirm the code.

• If the entered code doesn't exist the "Err"

message will be displayed for a few sec-
onds and then the message "Insert the
function code or press ALT - FNC to es-
cape" will recommence scrolling in the
upper LCD.

SETUP MODE

Advertising
This manual is related to the following products: