Hanna Instruments HI 93542 User Manual

Page 5

Advertising
background image

9

8

HI 93551R

HI 93552R

SPECIFICATIONS OF

HI 93551R & HI 93552R

Keyboard Functions:

ON/OFF : turn the meter ON and OFF.
T1/T2 : select the reading channel (T1, T2 or T1-T2).
HOLD : freeze the reading on display. The frozen value is also stored

in non-volatile memory.

CLR : clear the HI and LO values, reset the relative or average

measurement.

DISP : select Normal, Relative, Average or T1/T2 measuring mode.
ALT : enable the second function keys; the "ALT" tag turns on to

indicate that the second functions are enabled.

Note: The ALT key can be released before pressing the second

function key for 1-hand operation.

ALT/ CAL : press and hold for about 5 seconds to enter the

Calibration mode (with reading within ±3°C range).

(ALT +) AOFF : set auto-off delay (8 min, 60 min, disabled).
(ALT +) °C/°F : change reading units (°C or °F).
(ALT +) MR : recall memorized value.
(ALT +) KJT : select the thermocouple type.
(ALT +) BL : toggle the backlight ON and OFF.

SPECIFICATIONS

Range (*) K

-200.0 to 999.9°C / 1000 to 1371°C
-328.0 to 999.9°F / 1000 to 2500°F

J

-200.0 to 999.9°C

-328.0 to 999.9°F / 1000 to 1832°F

T

-200.0 to 400.0°C
-328.0 to 752.0°F

Resolution K

0.1°C (-149.9 to 999.9°C)

0.2°C (-200.0 to -150.0°C) / 1°C (1000 to 1371°C)

0.1°F (-24.9 to 999.9°F) / 0.2°F (-249.9 to -25.0°F)

0.3°F (-328.0 to -250.0°F) / 1°F (1000 to 2500°F)

J

0.1°C (-200.0 to 999.9°C)
0.1°F (-149.9 to 999.9°F)

0.2°F (-328.0 to -150.0°F) / 1°F (1000 to 1832°F)

T

0.1°C (-149.9 to 400.0°C)

0.2°C (-200.0 to -150.0°C)

0.1°F (0.0 to 752.0°F) / 0.2°F (-270.0 to -0.1°F)

0.3°F (-328.0 to -270.1°F)

Accuracy

±0.5°C (-100.0 to 999.9°C) / ±1°C (outside)

(@20°C/68°F) ±1°F (-148.0 to 999.9°F) / ±1.5°F (outside)

for one year, excluding probe error

Typical EMC Deviation

± 3°C / ± 6°F

Battery

3x1.5V AA (IEC LR6) batteries,

approx. 500 hours of continuous use (BL off)

Auto-off

user selectable: 8 min, 60 min, disabled

Serial Port

unidirectional, 8-bit data, 1200 baud

Communication

(see "Serial Communication" section)

Environment

-10 to 60°C (14 to 140°F); RH 100%

Dimensions

150 x 80 x 36 mm (5.9 x 3.1 x 1.4")

Weight

235 g (8.3 oz.)

(*)

Range may be limited by probe.

Advertising
This manual is related to the following products: