Hanna Instruments HI 93510 User Manual

Page 4

Advertising
background image

7

6

HI 93512

HI 93522

SPECIFICATIONS OF

HI 93512 & HI 93522

Main features:

• Two independent measurement channels, T1 and T2
• HI 762 probes series
• Available HI 762 accuracy test plugs
• Display of temperature in degrees Celsius or Fahrenheit
• Range -50 to 150°C or -58 to 302°F
• HOLD function
• HI and LO (Max and Min) temperature values always displayed

on the LCD

• DISP function to select the information to be displayed (Normal,

Relative or T1/T2)

• Remaining battery life indication / low battery detection
• Waterproof casing
• Auto-off capability (HI 93522 only)
• MR function for recalling memorized reading (HI 93522 only)
• Backlight feature (HI 93522 only)
• User calibration at 0°C (HI 93522 only)

SPECIFICATIONS

Range

-50.0 to 150.0°C
-58.0 to 302.0°F

Resolution

0.1°C

0.1°F (-58.0 to 230.0°F) / 0.2°F (outside)

Accuracy

±0.4°C / ±0.8°F

for one year, excluding probe error

Typical EMC

±0.5°C / ±1°F

Deviation

with HI 762BL probe

Battery

3x1.5V AA (IEC LR6) batteries,

approx. 2000 hours of continuous use (BL off)

Auto-off

user selectable: 8 min, 60 min, disabled

(HI 93522 only)
Environment

-10 to 50°C (14 to 122°F); RH 100%

Dimensions

150 x 80 x 36 mm (5.9 x 3.1 x 1.4")

Weight

235 g (8.3 oz.)

Keyboard Functions:

ON/OFF : turn the meter ON and OFF.
T1/T2 : select the reading channel (T1, T2 or T1-T2).
HOLD : freeze the reading on display. In HI 93522, the frozen value

is also stored in non-volatile memory.

CLR : clear the HI and LO values, reset the relative measurement.
DISP : select Normal, Relative or T1/T2 measuring mode.
ALT (HI 93522 only) : enable the second function keys; the "ALT" tag

turns on to indicate that the second functions are enabled.

Note: The ALT key can be released before pressing the second

function key for 1-hand operation.

ALT/ CAL : press and hold for about 5 seconds to enter the

Calibration mode (with reading within ±3°C range).

(ALT +) AOFF : set auto power off delay (8min, 60min, disabled).
(ALT +) °C/°F : change reading units (°C or °F).
(ALT +) MR : recall memorized value.
(ALT +) BL : toggle the backlight ON and OFF.

Advertising