38

DIAG

3.C If the recorder full scale (upper limit) reads too

low

3.C.1 Press DIAG. The display will show the last diagnostic

mode and the right-most numerical digit will blink.

3.C.2 Enter code 102 using the and keys.

3.C.3 Press DIAG; “r up” will be displayed and the analyzer

will execute this mode fine tuning upper value.

3.C.4 The reading on multimeter will slowly

increase. When the desired value is
reached, press DIAG again to store it
and quit the mode.

3.D If the recorder full scale (upper limit) reads too

high

3.D.1 Press DIAG. The display will show the last diagnostic

mode and the right-most numerical digit will blink.

DIAG

DIAG

DIAG