Turn the meter on by pressing the button,
all segments will be displayed. When the
display shows “Add”, “C.1” with “Press”
blinking, the meter is ready.

Using supplied syringe add 2 mL (1 mL
twice) of HI 739AS reagent to the cuvette
and use the pipette to fill the cuvette up to
the mark with HI 739BS reagent.

Replace the cap and invert several times to
mix.

Place the cuvette into the meter and close
the meter’s cap.

Press and hold the button until the timer is
displayed on the LCD. Alternatively wait for
one minute and then press the button. When
the display shows “Add”, “C.2” with “Press”
blinking the meter is zeroed.

Remove the cuvette from the meter and
unscrew the cap. Using the unused syringe,
add exactly 1.00 mL of sample.

Replace the cap and invert several times to
mix.

Place the cuvette back into the meter.

Press and hold the button until the timer is
displayed on the LCD. Alternatively wait for
one minute and then press the button.

The instrument directly displays the
concentration of fluoride in ppm.

The meter automatically turns off after 10
minutes.

• It is important that the sample does not contain any debris.
• Whenever the cuvette is placed into the measurement cell, it must be dry outside,

and completely free of fingerprints, oil and dirt. Wipe it thoroughly with HI 731318
or a lint-free cloth prior to insertion.

• Shaking the cuvette can generate bubbles, causing higher readings. To obtain

accurate measurements, remove bubbles by swirling or by gently tapping the
cuvette.

• Do not let the reacted sample stand for too long after reagent is added, as

accuracy will be affected.

• After the reading it is important to immediately discard the sample, otherwise

the glass might become permanently stained.

REAGENT SETS
HI 739-26

Liquid Reagent Set for 30 tests

OTHER ACCESSORIES
HI 739-11

Fluoride HR Certified Standard Kit

HI 740142

1 mL graduated syringe

HI 740144P

Pipette tip for syringes (10 pcs)

HI 740028

1.5V AAA batteries (4 pcs)

HI 731318

Cloth for wiping cuvettes (4 pcs)

HI 731321

Glass cuvettes (4 pcs)

HI 731225

Cuvette cap for checker HC (4 pcs)

HI 93703-50

Cuvette cleaning solution (230 mL).

Recommendations for Users

Before using this product, make sure that it is entirely suitable for your specific application and for the
environment in which it is used.
Operation of this instrument may cause unacceptable interferences to other electronic equipment, thus
requiring the operator to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the instrument’s EMC
performance.
To avoid damage or burns, do not put the instrument in microwave oven. For the safety of you and the
instrument do not use or store the instrument in hazardous environments.

For additional information, contact your

dealer or the nearest

Hanna Customer Service Center.

To find a Hanna Office in your area,

visit our web site:

w w w . h a n n a i n s t . c o m

Hanna Instruments

reserves the right to modify the design, construction or

appearance of its products without advance notice.

Battery management

To save the battery, the instrument shuts down after 10 minutes of non-use.
One fresh battery lasts for a minimum of 5000 measurements. When the battery is dead the
instrument will display “bAd” then ”bAt” for 1 second and then turns off.
To restart the instrument, the battery must be replaced with a new one.
To replace the instrument’s battery:
• Turn the instrument off by holding the button until the meter shuts off.
• Turn the instrument upside down and remove the battery cover with a screwdriver.

• Remove the battery from its location and replace it with a new one, inserting the

negative end first.

• Insert the battery cover and replace the screw with a screwdriver.

IST739 11 /11

Measurement procedure:

Accessories:

Tips for an accurate measurement

10 mL

2x