Installation, Specifications – Hanna Instruments HI 700 Series User Manual

Page 5

Advertising
background image

9

8

INSTALLATION

HI 700 and HI 710
series offer a multitude
of possibilities, from
single and dual set-
points to ON/OFF or
PID dosage, isolated
outputs with user-select-
able zoom, bi-directional
RS485, recorder outputs
in mAmps and Volts.

Use the 3-wire Pt 100
temperature sensor to
compensate for the
cable resistance and
have a precise auto-
matic temperature
compensation of the
measurements in long
distance applications.

See the diagram for a
recommended installa-
tion.

SPECIFICATIONS

Ranges

0.0 to 199.9

μS, 0 to 1999 μS

0.00 to 19.99 mS, 0.0 to 199.9 mS
0.0 to 100.0 ppm, 0 to 1000 ppm (HI 710* only)
0.00 to 10.00 ppt, 0.0 to 100.0 ppt (HI 710* only)
-10.0 to 100.0 °C

Resolution

0.1 μS, 1 μS, 0.01 mS, 0.1 mS
0.1 ppm, 1 ppm (HI 710 models only)
0.01 ppm, 0.1 ppm (HI 710 models only)
0.1 °C

Accuracy

±0.5 % F.S. (EC and TDS)

(@20°C/68°F)

±0.5°C (0 to 70°C), ±1°C (outside)

Typical EMC Deviation

±2 % F.S. (EC and TDS) / ±0.5 °C

EC Calibration

Automatic or manual at 1 point

Temperature Compensation

Automatic or manual, -10 to 100°C with adjustable
temperature coefficient from 0.00 to 10.00%/°C

TDS Factor (HI 710 only)

Adjustable from 0.00 to 1.00

Outputs

Analog: isolated 0-1 mA, 0-20 mA and 4-20 mA;
0-5 Vdc, 1-5 Vdc and 0-10 Vdc; or digital: RS485

Analog Input

4-20 mA

Setpoint Relay

2 contact outputs SPDT, 5A-250 Vac, 5A-30 Vdc
(resistive load). Fuse protected: 5A, 250V FUSE

Alarm Relay

Contact output SPDT, 5A-250 Vac, 5A-30 Vdc
(resistive load). Fuse protected: 5A, 250V FUSE

Power Supply

230 ±10% Vac or 115 ±10% Vac, 50/60 Hz

Power Consumption

15 VA

Over Current Protection

400 mA 250V FAST FUSE

Environment

0 to 50°C (32 to 122°F); RH max 95%

Installation Category

II

Enclosure

Single case ½ DIN

Dimensions

Panel cutout: 140 x 140 mm
Instrument: 144 x 144 x 170 mm

Weight

Approximately 1.6 kg. (3.5 lb.)

* Note: actual TDS range for HI 710 models depends on TDS factor set.

Advertising