OPERATIONAL GUIDE

POWER CONNECTION

Plug the 12 VDC adapter into the power supply socket.
Notes: • This instrument uses non volatile memory to retain the calibration

parameters and all other settings, even when unplugged.

• Make sure a fuse protects the main line.

PROBE CONNECTION

For EC/TDS or temperature measurements connect the probe to the 7-pin
connector. Make sure the probe sleeve is properly inserted.
INSTRUMENT START-UP

• Turn the instrument on by pressing the

ON/OFF switch located on

the rear panel.

• All LCD tags are displayed and a beep is sounded while the

instruments perform a self test.

• The instrument will display “

LoAd” message and “ ” blinking until

initialization is complete.

Notes: • The instrument starts in the same range and mode as it was

at power off.

• The RANGE key toggles between measurement ranges:

- EC, TDS and NaCl.

CONDUCTIVITY MEASUREMENTS

Connect the conductivity probe to the instrument.
• Press the

RANGE key to enter conductivity measurement range (EC).

• Submerse the probe into the solution to be tested. The

sleeve holes must be completely submersed. Tap the
probe repeatedly to remove any air bubbles that may
be trapped inside the sleeve.

• The conductivity value will be displayed on the primary

LCD and the temperature on the secondary LCD.

• If the reading is out of range, the full-scale value (200.0 for Mtc/Atc

mode or 500.0 for actual conductivity) will be displayed blinking.