Accessories – Hanna Instruments HI 21 Series User Manual

Page 32

Advertising
background image

63

62

ACCESSORIES

pH CALIBRATION SOLUTIONS
HI 7004M

pH 4.01 buffer solution, 230 mL bottle

HI 7004L

pH 4.01 buffer solution, 500 mL bottle

HI 7004/L

pH 4.01 buffer solution, 1 L bottle

HI 7007M

pH 7.01 buffer solution, 230 mL bottle

HI 7007L

pH 7.01 buffer solution, 500 mL bottle

HI 7007/L

pH 7.01 buffer solution, 1 L bottle

HI 7010M

pH 10.01 buffer solution, 230 mL bottle

HI 7010L

pH 10.01 buffer solution, 500 mL bottle

HI 7010/L

pH 10.01 buffer solution, 1 L bottle

ORP SOLUTIONS
HI 7020M

200-275 mV test solution, 230 mL bottle

HI 7020L

200-275 mV test solution, 500 mL bottle

HI 7091M

Reducing pretreatment solution, 230 mL bottle

HI 7091L

Reducing pretreatment solution, 500 mL bottle

HI 7092M

Oxidizing pretreatment solution, 230 mL bottle

HI 7092L

Oxidizing pretreatment solution, 500 mL bottle

ELECTRODE MAINTENANCE SOLUTIONS
HI 70300M

Storage solution, 230 mL bottle

HI 70300L

Storage solution, 500 mL bottle

HI 7061M

General cleaning solution, 230 mL bottle

HI 7061L

General cleaning solution, 500 mL bottle

HI 7073M

Protein cleaning solution, 230 mL bottle

HI 7073L

Protein cleaning solution, 500 mL bottle

HI 7074M

Inorganic cleaning solution, 230 mL bottle

HI 7074L

Inorganic cleaning solution, 500 mL bottle

HI 7077M

Oil & Fat cleaning solution, 230 mL bottle

HI 7077L

Oil & Fat cleaning solution, 500 mL bottle

HI 7071

3.5M KCl+AgCl electrolyte solution, 4x50 mL bottle, for single
junction electrodes

HI 7072

1M KNO

3

electrolyte solution, 4x50 mL bottle

HI 7082

3.5M KCl electrolyte solution, 4x50 mL bottle, for double junc-
tion electrodes

pH

mV

pH

mV

pH

mV

pH

mV

pH

mV

0

990

1

920

2

860

3

800

4

740

5

680

6

640

7

580

8

520

9

460

10

400

11

340

12

280

13

220

14

160

Reducing pretreatment: immerse the electrode for a few min-
utes in HI 7091.
Oxidizing pretreatment: immerse the electrode for a few min-
utes in HI 7092.
If the pretreatment is not performed, the electrode will take
significantly longer to respond.
As with pH electrodes, gel-filled redox electrodes are more
suitable for industrial applications due to lesser mainte-
nance requirements. However, if working with refillable
electrodes, the electrolyte level should not fall more than
2½ cm (1") below the fill hole and topped up if necessary.
Use HI 7071 refill solution for single junction and HI 7082
for double junction electrodes.
In the event that measurements are performed with solutions
containing sulfides or proteins, the cleaning of the diaphragm
of the reference electrode must be performed more often to
maintain the proper functioning of the ORP electrode. There-
fore, immerse it into HI 7020 and measure the response; the
obtained value should be within 200 and 275 mV.
After this functional test, it is suggested to wash the electrode
thoroughly with water and proceed to the oxidizing or reduc-
ing pretreatment before taking measurements.
When not in use, the electrode tip should be kept moist and
far from any type of mechanical stress which might cause
damage. This can be achieved by installing the electrode in
such a way that it is constantly in a well filled with the sample
(stream or tank). When not in use, store the electrode with a
few drops of HI 70300 storage solution in the protective
cap.

Note

With industrial applications, it is always recommended to
keep at least one spare electrode handy. When anomalies
are not resolved with a simple maintenance, change the elec-
trode to see if the problem is alleviated.