PROBE MAINTENANCE:
To improve the probe performance and pro-
long its life, it is recommended to clean it
regularly.
• Immerse the tip of the probe in HI7061

Cleaning Solution for one hour.

• If a more thorough cleaning is required,

brush the metal pins with very fine sand-
paper.

• After cleaning, rinse the probe with tap

water and recalibrate the meter.

• When not in use, clean the probe be-

fore storing it away.

CONNECTIONS:
The wires from the probe cable must be con-
nected to the process EC or TDS controller as
shown below. The connections are color coded
for an easy installation.

Note: once the probe is attached to the meter,
it is recommended to calibrate it before tak-
ing any measurement. Follow the calibration
procedure in the meter's manual.

IST7634-00 09/05

SPECIFICATIONS

(typical @25 ºC)

:

Cell Constant

1 cm

-1

NTC sensors

10 K

Materials

PP and AISI316

Temperature

-5 to 50 ºC

Max. pressure

3 bar