ERRORS AND WARNINGS

The instrument shows clear messages when erroneous condition appears. Messages are also displayed
when the obtained values are outside expected range. The beeper is playing a beep on errors.

a) on calibration

Light High:

There is too much light to perform a measurement.

Light Low:

There is not enough light to perform a measurement.

No Light:

The instrument cannot adjust the light level. Please

check that the tube holder does not contain any debris.

Cal:

Calibration was not performed. Follow the instructions of

the measurement procedure for calibrating the meter.

b) on sample reading

Under range:

A blinking “0.0” indicates that the sample

absorbs less light than the zero reference.