PRINCIPLE OF OPERATION

Absorption of Light is a typical phenomenon of interaction between electromagnetic radiation and
matter. When a light beam crosses a substance, some of the radiation may be absorbed by
atoms, molecules or crystal lattices.
If pure absorption occurs, the fraction of light absorbed depends both on the optical path length
through the matter and on the

physical

-chemical characteristics of the substance according to the

Lambert-Beer Law:

-log

I

/

I

o

=

ε

λ

c d

or

A

=

ε

λ

c d

Where:

-log

I

/

I

o

=

Absorbance (A)

I

o

=

intensity of incident light beam

I

=

intensity of light beam after absorption

ε

λ

=

molar extinction coefficient at wavelength

λ

c

=

molar concentration of the substance

d

=

optical path through the substance

Therefore, the concentration "c" can be calculated from the absorbance of the substance as the
other factors are known.
Photometric chemical analysis is based on the possibility to develop an absorbing compound
from a specific chemical reaction between sample and reagents. Given that the absorption of a
compound strictly depends on the wavelength of the incident light beam, a narrow spectral
bandwidth should be selected as well as a proper central wavelength to optimize measurements.
The optical system of Hanna's HI 96750 colorimeter is based on special LED and narrow-band
interference filters to guarantee both high performance and reliable results.