Hanna Instruments HI 96724C User Manual

Page 16

Advertising
background image

16

• After a few seconds, the display will show

“-0.0-”

. The meter is now zeroed and ready

for validation.

• Remove the cuvette.

• Place the CAL CHECK

TM

Standard HI 96724-11

Cuvette B into the cuvette holder and ensure
that the notch on the cap is positioned securely
into the groove.

• Press CAL CHECK

TM

and the lamp, cuvette

and detector icons together with “CAL CHECK”
will appear on the display, depending on the
measurement phase.

• At the end of the measurement the display

will show the validation standard value.

The reading should be within specifications as reported in the CAL CHECK

TM

Standard

Certificate. If the value is found out of the specifications, please check that the cuvettes are free
of fingerprints, oil or dirt and repeat validation. If results are still found out of specifications,
then recalibrate the instrument.

• Press ZERO/CFM and the lamp, cuvette

and detector icons will appear on the
display, depending on the measurement
phase.

• Place the CAL CHECK

TM

Standard

HI 96724-11

Cuvette A into the cuvette

holder and ensure that the notch on the
cap is positioned securely into the groove.

Advertising