Validation procedure – Hanna Instruments HI 96717C User Manual

Page 8

Advertising
background image

14

15

• Press ZERO/CFM and the lamp, cuvette

and detector icons will appear on the
display, depending on the measurement
phase.

• After a few seconds, the display will show

“-0.0-”

. The meter is now zeroed and

ready for measurement.

• Remove the cuvette.

• Add 10 drops of HI 93717A-0 Phosphate

reagent A to the cuvette.

• Add the content of one packet of

HI 93717B-0

Phosphate reagent B to

the cuvette. Replace the cap and shake
gently until dissolution is complete.

• Replace the cuvette into the cuvette holder

and ensure that the notch on the cap is
positioned securely into the groove.

• Press and hold READ/

READ/

READ/

READ/

READ/ /TIMER

/TIMER

/TIMER

/TIMER

/TIMER

for three

seconds.

The display will show the

countdown prior to measurement. The
beeper is playing a beep at the end of
countdown period.

Alternatively, wait for 5 minutes and just
press READ/

READ/

READ/

READ/

READ/ /TIMER

/TIMER

/TIMER

/TIMER

/TIMER.

In both cases, the lamp, cuvette and
detector icons will appear on the display,
depending on the measurement phase.

• At the end of measurement, the instrument

directly displays concentration in mg/L of
phosphate on the LCD.

INTERFERENCES
Calcium (Ca

2+

):

above 10,000 mg/L as CaCO

3

Chloride (Cl

-

):

above 150,000 mg/L as Cl

-

Ferrous iron (Fe

2+

):

above 100 mg/L as Fe

2+

Magnesium (Mg

2+

):

above 40,000 mg/L as CaCO

3

Sulfide (S

2+

)

VALIDATION PROCEDURE

Use the validation procedure to ensure that the instrument is properly calibrated.
Warning:

Do not validate the instrument with any standard solutions other than the HANNA CAL

CHECK

TM

Standards, otherwise erroneous results will be obtained.

Turn the meter on by pressing ON/OFF.

• Place the CAL CHECK

TM

Standard

HI 96717-11

Cuvette A into the cuvette

holder and ensure that the notch on the
cap is positioned securely into the groove.

• When the beeper sounds briefly and the

LCD displays dashes, the meter is ready.

Advertising