Hanna Instruments HI 96717 User Manual

Hanna Instruments Measuring instruments

Advertising
background image

1. GLP/ key: press to enter

GLP mode. In calibration mode press to

edit the date and time.

2. CAL CHECK key: press to perform the validation of the meter, or

press and hold for three seconds to enter

calibration mode.

3. ZERO/CFM key: press to zero the meter prior to measurement, to

confirm edited values or to confirm factory calibration restore.

4. READ/ /TIMER key: In

measurement mode, press to make a

measurement, or press and hold for three seconds to start a pre-
programmed countdown prior to measurement. In

GLP mode press to

view the next screen.

5. ON/OFF key: to turn the meter on and off.

6. Liquid Crystal Display (LCD)
7. Cuvette alignment indicator

8. Cuvette holder

3

DISPLAY ELEMENTS DESCRIPTION:

INSTRUCTION MANUAL

HI 96717

HI 96717

HI 96717

HI 96717

HI 96717

Phosphate High

Range ISM

w w w . h a n n a i n s t . c o m

Preliminary examination:

Please examine this product carefully. Make sure that the instrument is
not damaged. If any damage occured during shipment, please notify
your Dealer.
Each HI 96717 Ion Selective Meter is supplied complete with:
• Two Sample Cuvettes and Caps
• 9V Battery
• Instruction Manual

Note:

save all packing material until you are sure that the instrument
works correctly. Any defective item must be returned in its original
packing.

For more details about spare parts and
accessories see “Accessories”

Dear Customer,
Thank you for choosing a Hanna product. This manual will provide you
with the necessary information for the correct use of the instrument. Please
read it carefully before using the meter. If you need additional technical
information, do not hesitate to e-mail us at [email protected].

Technical specifications:

Range

0.0 to 30.0 mg/L

Resolution

0.1 mg/L

Accuracy

±1 mg/L ±4% of reading @ 25°C

Typical EMC Dev.

±0.2 mg/L

Light Source

Tungsten lamp

Light Detector

Silicon Photocell with narrow band interfer-
ence filter @ 525 nm

Method

Amino Acid Method, adapted from

Standard

Method for the Examination of Water and
Wastewater.

Environment

0 to 50°C (32 to 122°F);
max 95% RH non-condensing

Battery Type

1 x 9 volt

Auto-Shut off

After 10' of non-use in

measurement mode;

after 1 hour of non-use in

calibration mode;

with last reading reminder.

Dimensions

192 x 102 x 67 mm (7.6 x 4 x 2.6”)

Weight

290 g (10 oz.).

Functional description:

Light High:

There is too much light to perform

a measurement. Please check the preparation of
the zero cuvette.

Light Low:

There is not enough light to perform

a measurement. Please check the preparation of
the zero cuvette.

No Light:

The instrument cannot adjust the

light level. Please check that the sample does
not contain any debris.

Inverted cuvettes:

The sample and the zero

cuvette are inverted.

Zero:

A zero reading was not taken. Follow the

instructions of the measurement procedure for
zeroing the meter.

Under range:

A blinking “0.0” indicates that

the sample absorbs less light than the zero
reference. Check the procedure and make sure
you use the same cuvette for reference (zero)
and measurement.

Over Range:

A flashing value of the maximum

concentration indicates an over range condition.
The concentration of the sample is beyond the
programmed range: dilute the sample and re-
run the test.

Standard Low:

The standard reading is less

than expected.

Standard High:

The standard reading is higher

than expected.

Errors and warnings:

Measurement procedure:

1• Turn the meter on by pressing ON/OFF.
2• When the beeper sounds briefly and the

LCD displays dashes, the meter is ready.
The blinking “ZERO” indicates that the
instrument needs to be zeroed first.

3• Fill the cuvette with 10 mL of unreacted

sample, up to the mark, and replace the cap.

4• Place the cuvette into the cuvette holder

and ensure that the notch on the cap is
positioned securely into the groove.

5• Press ZERO/CFM and the lamp, cuvette

and detector icons will appear on the display,
depending on the measurement phase.

6• After a few seconds the display will show

“-0.0-”. The meter is now zeroed and
ready for measurement.

7• Remove the cuvette.
8• Add 10 drops of HI 93717A-0 Phosphate

Reagent A to the cuvette.

9• Add the content of one packet of HI 93717B-0

Phosphate Reagent B to the cuvette.

10•Replace the cap and shake gently until

dissolution in complete.

11•Replace the cuvette into the cuvette

holder and ensure that the notch on the
cap is positioned securely into the groove.

12•Press and hold READ/ /TIMER

for three

seconds.

The display will show the

countdown prior to measurement. The
beeper is playing a beep at the end of
countdown period.
Alternatively, wait for 5 minutes and just
press READ/ /TIMER.
In both cases, the lamp, cuvette and
detector icons will appear on the display,
depending on the measurement phase.

13•The instrument directly displays concentration

in mg/L of phosphate on the LCD.

Measurement

1. The measuring scheme (lamp, cuvette, detector), appears during

different phases of zero or reading measurement

2. Error messages and warnings
3. The baterry icon shows the charging level of the batery
4. The hourglass appears when an internal checking is in progress
5. Status messages
6. The chronometer appears when the reaction timer is running
7. The month, day and date icons appear when a date is displayed
8. Four digit main display
9. Measuring units
10. Four digit secondary display

ON ZERO READING:

ON SAMPLE READING:

DURING CALIBRATION PROCEDURE:

OTHER ERRORS AND WARNINGS:

Cooling lamp:

The instrument waits for the

lamp to cool down.

Battery low:

The battery must be replaced

soon.

Dead battery:

This indicates that the battery

is dead and must be replaced. Once this
indication is displayed, the meter will lock up.
Change the battery and restart the meter.

Cap error:

Appears when external light enters

in the analysis cell. Assure that the cuvette cap
is present.

2

4, 11

5-6

8

10

9