KEYPAD DESCRIPTION
• ON/OFF: to turn the meter on and off.
• ZERO/CFM: this is a bi-functional key. Just press to zero the meter prior to measurement,

or to confirm edited values. In

calibration mode

press to confirm factory calibration restore.

• READ/ /TIMER: this is a multi-functional key. In

measurement mode

, press to make a

measurement, or press and hold for three seconds to start a pre-programmed countdown prior
to measurement. In

GLP mode

press to view the next screen.

• CAL CHECK: this is a bi-functional key. Just press to perform the validation of the meter, or

press and hold for three seconds to enter

calibration mode

.

• GLP/ : this is a bi-functional key. Just press to enter

GLP mode

. In

calibration mode

press

to edit the date and time.

OPERATING MODES

Measurement mode:

default operation mode, enables both validation and measurement.

Calibration mode:

may be entered

by keeping CAL CHECK pressed

for three seconds (

the

“CAL”

tag appears), it enables calibration of the instrument.

GLP mode

may be entered by pressing GLP/ (“GLP” appears), it enables consulting of

user calibration date or restore factory calibration.

DISPLAY ELEMENTS DESCRIPTION

1) The measuring scheme (lamp, cuvette, detector), appears during different phases of zero or

reading measurement

2) Error messages and warnings
3) The battery icon indicates the charge state of the battery
4) The hourglass appears when an internal check is in progress

5) Status messages
6) The chronometer appears when the reaction timer is running
7) The month, day and date icons appear when a date is displayed

8) Four digit main display

9) Measuring units

10) Four digit secondary display