CALIBRATION PROCEDURE
• Turn the meter on by pressing ON/OFF.
• When the LCD displays “715”, it is ready.

• Enter the calibration mode by holding CAL CHECK

for three seconds

.

• The date of the last calibration appears (e.g.:

month “01”, day “08”). “F.CAL” means that
the factory calibration is selected.
Note: at this point it is possible to reset the
instrument to restore FACTORY CALIBRATION,
see instructions page 18.

• Place the CAL CHECK Standard Cuvet A into the

holder and ensure that the notch on the cap is
positioned securely into the groove.

• Press ZERO and “SIP” will blink on the display.

• After a few seconds the display will show

“-0.0-”. The meter is now zeroed and ready for
calibration.

• Remove the cuvet.

• Place the CAL CHECK Standard HI 95715-11

Cuvet B into the holder and ensure that the
notch on the cap is positioned securely into the
groove.

CALIBRATION PROCEDURE

Note: it is possible to interrupt calibration procedure at any time by pressing ON/OFF.
Warning: do not validate or calibrate the instrument with standard solutions other than the Hanna
CAL CHECK Standards, otherwise erroneous results will be obtained. For accurate Validation and
Calibration, please perform test at room temperature: 18 to 25°C (64.5 to 77.0°F).