Hanna Instruments HI 95707 User Manual

Page 5

Advertising
background image

5

16

Range

0.000 to 0.600 mg/L (as NO

2

-N)

Resolution

0.001 mg/L

Precision

±0.001 mg/L @ 0.100 mg/L

Typical EMC Deviation

±0.001 mg/L

Light Source

Tungsten lamp with narrow band interference filter @ 525 nm

Light Detector

Silicon Photocell

Method

Adaptation of an EPA approved method. The reaction between Nitrite

and the reagent causes a pink tint in the sample.

Environment

0 to 50°C (32 to 122°F); max 95% RH non-condensing

Battery Type

1 x 9 volt

Auto-Shut off

After 10' of non-use in measurement mode;
after 1 hour of non-use in calibration mode.

Dimensions

180 x 83 x 46 mm (7.1 x 3.3 x 1.8")

Weight

290 g (10 oz.)

REQUIRED REAGENTS

Code

Description

Quantity/test

HI 93707-0

Nitrite Low Range Reagent

1 packet

SPECIFICATIONS

Precision is how closely repeated

measurements agree with each other.

Precision is usually expressed as standard

deviation (SD).
Accuracy is defined as the nearness of a test

result to the true value.
Although good precision suggests good

accuracy, precise results can be inaccurate. The

figure explains these definitions.
In a laboratory using a standard solution of

0.100 mg/L Nitrogen-Nitrite and a

representative lot of reagent, an operator

obtained with a single instrument a standard

deviation of 0.001 mg/L.

SPECIFICATIONS

PRECISION AND ACCURACY

CALIBRATION PROCEDURE
• Turn the meter on by pressing ON/OFF.

When the LCD displays “707”, it is ready.

• Enter the calibration mode by holding CAL CHECK

for three seconds.

• The date of the last calibration appears (e.g.:

month “01”, day “08”). “F.CAL” means that the
factory calibration is selected.
Note: at this point it is possible to reset the
instrument to restore FACTORY CALIBRATION, see
instructions page 18.

• Place the CAL CHECK™ Standard Cuvet A into

the holder and ensure that the notch on the cap is
positioned securely into the groove.

• Press ZERO and “SIP” will blink on the display.

• After a few seconds the display will show

“-0.0-”. The meter is now zeroed and ready for
calibration.

• Remove the cuvet.

• Place the CAL CHECK™ Standard HI 95707-11

Cuvet B into the holder and ensure that the notch

on the cap is positioned securely into the groove.

Note: it is possible to interrupt calibration procedure at any time by pressing ON/OFF.
Warning: do not validate or calibrate the instrument with standard solutions other than the Hanna CAL
CHECK™ Standards, otherwise erroneous results will be obtained. For accurate Validation and Calibration
results, please perform test at room temperature: 18 to 25°C (64.5 to 77.0°F).

or

CALIBRATION PROCEDURE