HI 93743

Iron LR & pH

The HI 93743 meter measures the pH and the Iron (Fe) content in
water and wastewater.

SPECIFICATIONS

Range

Fe 0 to 400 µg/L

pH 5.9 to 8.5

Resolution

Fe 1 µg/L

pH 0.1

Accuracy

Fe ±10 µg/L ±8% of reading

pH ±0.1

Typical EMC

Fe ±1 µg/L

Deviation

pH ±0.1

Light Source

Light Emitting Diode @ 555 nm

Method

For Iron, Adaptation of the TPTZ Method. The

reaction between iron and the reagent causes a

blue tint in the sample.

For pH, Phenol red method. The reaction be-

tween pH and the reagent causes a red tint in

the sample.

REQUIRED REAGENTS

Code

Unit

Description

Quantity

HI 93746-0

Iron LR

TPTZ Reagent

2 packets

HI 93710-0

pH

Phenol red

5 drops