Instruction Manual

HI 93731

Zinc ISM

HI 93731 is warranted for two years against defects in

workmanship and materials when used for its intended purpose and

maintained according to instructions.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed

maintenance are not covered.
If service is required, contact the dealer from whom you purchased

the instrument. If under warranty, report the model number, date of

purchase, serial number and the nature of the failure. If the repair

is not covered by the warranty, you will be notified of the charges

incurred. If the instrument is to be returned to Hanna Instruments,

first obtain a Returned Goods Authorization Number from the

Customer Service department and then send it with shipment costs

prepaid. When shipping any instrument, make sure it is properly

packaged for complete protection.
To validate your warranty, fill out and return the enclosed warranty

card within 14 days from the date of purchase.
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner, Hanna Instruments
Inc., Woonsocket, Rhode Island, 02895 , USA.

Dear Customer,
Thank you for choosing a Hanna product. This manual will

provide you with the necessary information for the correct

operation of the meter. Please read it carefully before using

the meter. If you need additional technical information, do

not hesitate to e-mail us at [email protected].
This instrument is in compliance with directives.

Remove the instrument from the packing material and examine

it carefully to make sure that no damage has occurred during

shipment. If there is any damage, notify your Dealer.
Each Ion Specific Meter is supplied complete with

• 9V Battery

• Two Sample Cuvets and Caps

• One Transport Cap

Note: Conserve all packing material until the instrument has

been observed to function correctly. Any defective item

must be returned in its original packing.

The HI 93731 meter measures the zinc content in water and
wastewater in the 0.00 to 3.00 mg/L (ppm) range.
The meter uses an exclusive positive-locking system to ensure

that the cuvet is in the same place every time it is placed into

the measurement cell.
The reagents are in liquid and powder form and are supplied

in bottles and in packets. The amount of reagent is precisely

dosed to ensure maximum repeatability.
Display codes aid the user in routine operations.
The meters have an auto-shut off feature that will turn the

instrument off after 10 minutes of non-use.

SPECIFICATIONS

Range

0.00 to 3.00 mg/L

Resolution

0.01 mg/L

Accuracy

±0.03 mg/L ±3% of reading

Typical EMC

±0.01 mg/L

Deviation
Light Source

Light Emitting Diode @ 585 nm

Method

Adaptation of the Standard Methods
for the Examination of Water and
Wastewater, 20

th

edition,

Zincon

method. The reaction between zinc and

the reagent causes a brownish-green to

a blue tint in the sample

Light Detector

Silicon Photocell

Environment

0 to 50°C (32 to 122°F);
max 95% RH non-condensing

Battery Type/Life 1 x 9 volt/40 hours

Auto-Shut off

After 10' of non-use

Dimensions

180 x 83 x 46 mm

(7.1 x 3.3 x 1.8")

Weight

290 g (10 oz.).

REQUIRED REAGENTS

Code

Description

Quantity

HI 93731A-0

Reagent A

1 packet

HI 93731B-0

Reagent B Cyclohexanone

0.5 mL

REAGENT SETS

HI 93731-01

Reagents for 100 tests

HI 93731-03

Reagents for 300 tests

HANNA instruments®

reserves the right to modify the design,

construction and appearance of its products without advance notice.

This Instrument is in

Compliance with the CE Directives

www.hannainst.com

Warranty

General Description

Preliminary Examination

Specifications

IST93731R2

01/07

This indicates that the meter is in a
ready state and zeroing can be
performed.
Sampling in Progress. This prompt
appears each time the meter is
performing a measurement.
This indicates that the meter is in a
zeroed state and measurement can be
performed.
A zero reading was not taken. Insert
a sample before adding reagent and
press ZERO.
Under range. A blinking "0.00"
indicates that the sample absorbs less
light than the zero reference. Check
the procedure and make sure you use
the same cuvet for reference (zero)
and measurement.
Over range. A flashing value higher
than the maximum concentration
readable (see specifications) indicates
that the sample absorbs too much
light, meaning that the concentration
is too high. Dilute the sample.
Light over range. The cuvet is not
inserted correctly and an excess
ambient light is reaching the detector.
If the cover is properly installed, then
contact your dealer or the nearest
Hanna Customer Service Center.
Light under range. The zero sample is
too dark for proper zeroing. If this is
not the case, contact your dealer or
the nearest Hanna Customer Service
Center.
The "V" indicates that the battery
voltage is getting low and the battery
needs to be replaced.
This indicates that the battery is dead
and must be replaced.
Note: once this indication is
displayed, the meter will lockup.
Change the battery to restart.

Display Code Guide

Battery replacement must only take
place in a non-hazardous area
using a 9V alkaline battery.
Simply slide off the battery cover
on the back of the meter. Detach
the battery from the terminals and
attach a fresh 9V battery while
paying attention to the correct
polarity. Replace the battery and
the cover.

9V

SLIDE

OFF

Battery Replacement