Hanna Instruments HI 93712 User Manual

Page 2

Advertising
background image

ON

OFF

• Turn the meter on by

pressing ON/OFF.

• When the LCD displays

"- - -", it is ready.

• Fill a graduated beaker

with 50 mL of sample.

• Add the contents of one

packet of HI 93712A-0
Aluminum Reagent A and
mix until dissolution is
complete.

• Add the content of one

packet of HI 93712B-0
Aluminum Reagent B and
mix until dissolution is

dissolution is

dissolution is

dissolution is

dissolution is

complete

complete

complete

complete

complete. This is the
sample.

• Fill two cuvets up to 1.5

cm (¾") below the rim
with 10 mL of sample each.

• Add the content of one HI

93712C-0 Aluminum
Reagent C to one only of
the two cuvets.

• Replace the cap and shake

gently. This is the blank.

ZERO

READ

TIMED

• Place the blank into the

holder and ensure that the
notch on the cap is
positioned securely into the
groove.

READ

DIRECT

9V

SLIDE

OFF

REAGENT SETS

HI 93712-01 Reagents for 100 tests Aluminum
HI 93712-03 Reagents for 300 tests Aluminum

OTHER ACCESSORIES

HI 710009

Blue rubber boot

HI 710010

Orange rubber boot

HI 721310

9V battery (10 pcs)

HI 731318

Tissue for wiping cuvets (4 pcs)

HI 731321

Glass cuvets (4 pcs)

HI 731325

Caps for cuvets (4 pcs)

HI 93703-50 Cuvets cleaning solution (230 mL).

Operational Guide

MEASUREMENT PROCEDURE

The instruction listed below should be carefully followed
during testing to ensure best accuracy.
• Do not touch the cuvet walls with hands.
• In order to maintain the same conditions during the

zeroing and the measuring phases, it is necessary to close
the cuvet to prevent any contamination.

• Do not let the test sample stand too long after reagent is

added or accuracy will be lost.

• Whenever the cuvet is placed into the measurement cell,

it must be completely free of fingerprints, oil or dirt.
Wipe it thoroughly with HI 731318 or a lint-free cloth
prior to insertion.

• It is important that the sample does not contain any

debris. This would corrupt the readings.

• It is possible to take multiple readings in a row, but it is

recommended that a zero reading be taken for each
sample and that the same cuvet is used for zeroing and
measurement.

• It is important to discard the sample immediately after

the reading is taken because the glass might become
permanently stained.

• Shaking the cuvet can generate bubbles in the sample,

causing higher readings. To obtain accurate measurements,
remove such bubbles by swirling or by gently tapping the
vial.

• All the reaction times reported in this manual are referred

to 20°C (68°F). As a general rule of thumb, they should
be doubled at 10°C (50°F) and halved at 30°C (86°F).

Battery replacement must only take place in a non-hazardous
area using a 9V alkaline battery.
Simply slide off the battery cover on the back of the meter.
Detach the battery from the terminals and attach a fresh 9V
battery while paying attention to the correct polarity. Replace
the battery and the cover.

Tips for an Accurate Measurement

Battery Replacement

Accessories

Recommendations for Users
Before using these products, make sure that they are entirely suitable for the environment
in which they are used.
Operation of these instruments in residential area could cause unacceptable interferences
to radio and TV equipments, requiring the operator to take all necessary steps to correct
interferences.
Any variation introduced by the user to the supplied equipment may degrade the
instruments' EMC performance.
To avoid damages or burns, do not perform any measurement in microwave ovens.

CE Declaration of Conformity

Read the relevant SDS sheets before performing the test.

Safety Data Sheets are available at: www.hannainst.com

Safety Data Sheets

# 1

# 2

# 1

• Press READ TIMED and the display will show the countdown

prior to zeroing the blank. Alternatively wait for 15
minutes and press ZERO. In both cases “SIP” will appear
during zeroing.

• Wait for a few seconds and

the display will show
"-0.0-". Now the meter is
zeroed and ready for
measurement.

• Remove the blank and insert

the other cuvet into the
instrument.

# 2

• Press READ DIRECT. "SIP" will appear during

measurement.

• The instrument directly displays concentration in mg/L of

aluminum on the Liquid Crystal Display.

INTERFERENCES

Interference may be caused by:
Iron above 20 mg/L
Alkalinity above 1000 mg/L
Phosphate above 50 mg/L
Fluoride will interfere at all levels.