Hanna Instruments HI 83225 User Manual

Page 32

Advertising
background image

32

SPECIFICATIONS

Range

0 to 150 mg/L

Resolution

2.5 mg/L

Accuracy

±5 mg/L ±8% of reading at 25 °C

Typical EMC Dev. ±2.5 mg/L
Light Source

Tungsten lamp with narrow band interference filter @ 525 nm

Method

Adaptation of the cadmium reduction method. The reaction between nitrate-nitrogen
and the reagent causes an amber tint in the sample.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93728-0

Powder reagent 1 packet

REAGENT SETS

HI 93728-01 Reagents for 100 tests
HI 93728-03 Reagents for 300 tests
For other accessories see page 57.

MEASUREMENT PROCEDURE

Note: for sample preparation follow the NUTRIENTS SOLUTION (MR) procedure at page 21.
• Select the

Nitrate MR method using the procedure described in

the

Method Selection section (see page 12).

• Using the pipette, fill the cuvette with 6 ml of sample, up to

half of its height, and replace the cap.

• Place the cuvette into the holder and close the lid.

• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for

measurement.

• Remove the cuvette and add the content of one packet of

HI 93728-0 reagent.

NITRATE MEDIUM RANGE

Nitrate MR

6 mL