• Remove the cuvette and add the content of one packet of

HI 93723-0 reagent. Replace the cap and shake vigorously
for about 10 seconds.

• Reinsert the cuvette into the instrument.

• Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait

for 6 minutes and press Read. When the timer ends the meter will perform the reading. The
instrument displays concentration in μμμμμg/L of chromium VI.

• Press  or  to access the second level functions.
• Press the Chem Frm key to convert the result in μg/L of Chromate (CrO

4

2

¯) and Dichromate (Cr

2

O

7

2

¯).

• Press  or  to return to the measurement screen.

INTERFERENCES

Interference may be caused by:
Vanadium above 1 ppm. However, waiting 10 minutes before reading, the interference is removed.
Iron above 1 ppm.
Mercurous and mercuric ions cause slight inhibition of the reaction.

Chromium VI HR