Tips for an accurate measurement – Hanna Instruments HI 83218 User Manual

Page 7

Advertising
background image

7

KEYPAD DESCRIPTION

The keypad contains 8 direct keys and 3 functional keys with the following functions:

Press to perform the function displayed above it on the LCD.

ESC

Press to exit the current screen.

Press to access the select method menu.

Press to move up in a menu or a help screen, to increment a set value, to access second level
functions.

Press to move down in a menu or a help screen, to decrement a set value, to access second
level functions.

Press to log the current reading.

RCL

Press to recall the log.

HELP

Press to display the help screen.

SETUP

Press to access the setup screen.

TIPS FOR AN ACCURATE MEASUREMENT

The instructions listed below should be carefully followed during testing to ensure most accurate results.
• Color or suspended matter in large amounts may cause interference, and should be removed by

treatment with active carbon and filtration: refere to Sample Preparation Chapter (page 17).

• Ensure the cuvette is filled correctly: the liquid in the cuvette forms a convexity on the top; the bottom

of this convexity must be at the same level as the 10 mL mark.

COLLECTING AND MEASURING SAMPLES

• In order to measure exactly 0.5 mL of reagent with the 1 mL syringe:

(a) push the plunger completely into the syringe and insert the tip into the solution.
(b) pull the plunger up until the lower edge of the seal is exactly on the 0.0 mL mark.