Hanna Instruments HI 83214 User Manual

Page 43

Advertising
background image

43

OXYGEN DEMAND, CHEMICAL MEDIUM RANGE

SPECIFICATIONS

Range

0 to 1500 mg/L COD

Resolution

1 mg/L

Accuracy

±15 mg/L or ±4 % of reading @ 25 °C, whichever is greater

Light Source

Tungsten lamp with narrow band interference filter @ 610 nm

Method

Adaptation of the USEPA 410.4 approved method for the COD determination on surface
waters and wastewaters. Oxidizable organic compounds reduce the dichromate ion
(orange) to the chromic ion (green). The amount of chromic ion formed is determined.

REQUIRED REAGENTS

Code

Description

Q.ty/test

Q.ty/set

COD-MR*

Reagent Vial

1 vial

25 vials

DEIONIZED120

Deionized Water

2.0 mL

optional

*

Reagent Vial identification: COD B, white label.

Note: Store the unused vials in their container in a cool and dark place.

REAGENT SET

HI 93754B-25 Reagents for up to 24 tests

REQUIRED ACCESSORIES

HI 839800-01 Hanna reactor (115 VAC)
HI 839800-02 Hanna reactor (230 VAC)
HI 740216

Test tube cooling rack (25 holes)

HI 740217

Laboratory bench safety shield

For other accessories see page 68.

MEASUREMENT PROCEDURE

Before using the reagent kit carefully read all the instructions and the Manual Safety Data Sheet

(MSDS). Pay particular attention to all warnings, cautions and notes. Failure to do so may result
in serious injury to the operator.

Reagent Blank Correction: This method requires a reagent blank correction. A single blank vial may be used
more than once. The blank vial is stable for several months (room temperature). For most accurate measurement,
run a blank for each set of measurements and always use the same lot of reagents for blank and samples.

COD MR

Advertising