• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for

measurement.

• Fill a second cuvette up to the mark with 10 mL of the reacted sample.

• Add 3 drops of Dispersing Agent (HI 93703-51),

replace the cap and invert gently to mix for about
10 seconds.

• Insert the second cuvette into the instrument.

• Press Read to start the reading. The instrument displays the results in mg/L of silver.

INTERFERENCES

Interference may be caused by:
Al

3+

above 30 mg/L

Fe

2+

above 1.5 mg/L

Ca

2+

above 1000 mg/L as CaCO

3

Fe

3+

above 10 mg/L

Cd

2+

above 20 mg/L

K

+

above 500 mg/L

Cl¯ above 8000 mg/L

Mn

2+

above 25 mg/L

Co

2+

above 1.5 mg/L

Mg

2+

above 1000 mg/L as CaCO

3

Cr

3+

above 20 mg/L

Na

+

above 5000 mg/L

Cr

6+

above 40 mg/L

Ni

2+

above 1.5 mg/L

Cu

2+

above 15 mg/L

Pb

2+

above 20 mg/L

F¯ above 20 mg/L

Zn

2+

above 30 mg/L