10

11

Pay attention to push the cuvette completely down in the holder.

• In order to avoid reagent leaking and to obtain more accurate measure-

ments, close the cuvette first with the supplied HDPE plastic stopper
and then the black cap.

• Whenever the cuvette is placed into the measurement cell, it must be dry

outside, and free of fingerprints, oil or dirt. Wipe it thoroughly with
HI 731318 or a lint-free cloth prior to insertion.

• Shaking the cuvette can generate bubbles in the sample, causing higher

readings. To obtain accurate measurements, remove such bubbles by
swirling or by gently tapping the cuvette.

• Do not let the reacted sample stand too long after reagent is added, or

accuracy will be lost.

• It is possible to take multiple readings in a row, but it is recommended to

take a new zero reading for each sample and to use the same cuvette for
zeroing and measurement when possible (for most precise results follow the

measurement procedures carefully).

• Discard the sample immediately after the reading is taken, or the glass

might become permanently stained.

• All the reaction times reported in this manual are at 20°C (68°F). As a general rule of thumb, they

should be doubled at 10°C (50°F) and halved at 30°C (86°F).

REAGENT BLANK CORRECTION

• Some methods require a “reagent blank correction”. The blank and the sample are prepared exactly

in the same way, only the blank is deionized water instead of sample.
A blank cuvette may be used more than once: stability and storing conditions are described for each
method in the related chapter.

INTERFERENCES

• In the method measurement section the most common interferences that may be present in an average

wastewater matrix have been reported. It may be that for a particular treatment process other

compounds do interfere with the method of analysis.

METHOD REFERENCE TABLE

HI 83208 - WATER CONDITIONING

Method

Method

Page

description

1

Ammonia MR

17

2

Ammonia LR

19

3

Free Chlorine

21

4

Total Chlorine

24

5

Copper HR

27

6

Copper LR

29

7

Fluoride

31

8

Iron HR

33

9

Iron LR

35

10

Manganese HR

38

11

Manganese LR

40

12

Molybdenum

43

Method

Method

Page

description

13

Nickel HR

46

14

Nickel LR

48

15

Nitrate

51

16

Dissolved Oxygen

53

17

pH

55

18

Phosphate HR

57

19

Phosphate LR

59

20

Phosphorus

61

21

Silica

63

22

Silver

66

23

Zinc

69

HEALTH & SAFETY

The chemicals contained in the reagent kits may be hazardous if improperly handled. Read the
Material Safety Data Sheet (MSDS) before performing tests.

Safety equipment: Wear suitable eye protection and clothing when required, and follow instructions carefully.
Reagent spills: If a reagent spill occurs, wipe up immediately and rinse with plenty of water.
If reagent contacts skin, rinse the affected area thoroughly with water. Avoid breathing released vapors.

Advertising