The instrument shows clear warning messages when erroneous conditions appear and when measured values are
outside the expected range. These messages are described below.

No Light: The light source is not functioning properly.

Light Leak: There is an excess amount of ambient light reaching the

detector.

Inverted cuvettes: The sample and the zero cuvettes are inverted.

Battery Low: The battery capacity is lower than 10%.

Light Low: The instrument cannot adjust the light level. Please check

that the sample does not contain any debris.

ERRORS AND WARNINGS

Light High: There is too much light to perform a measurement. Please

check the preparation of the zero cuvette.

INTERFERENCES

Interference may be caused by:

Aluminum above 6 mg/L

Cadmium above 0.5 mg/L

Copper above 5 mg/L

Iron above 7 mg/L

Manganese above 5 mg/L

Nickel above 5 mg/L

Zinc

• Place the cap and insert the cuvette into the instrument

and close the lid.

• Press ZERO key. The display will show “-0.0-” when the

meter is zeroed and ready for measurement.

• Remove the cuvette and add 0.5 mL of HI 93731B-0

cyclohexanone to the cuvette.
Note: To prevent any contamination from the polycarbon-

ate cap, prior to replacing it, close the sample cuvette
with the supplied HDPE plastic stopper.

• Replace the cap and mix the sample for 15 seconds.

• Insert the sample into the instrument.

• Press TIMER and the display will show the countdown

prior to the measurement or, alternatively, wait for 3

minutes and 30 seconds and press READ. When the timer
ends the meter will perform the reading.

• The instrument displays the results in mg/L of zinc.