Hanna Instruments HI 83203 User Manual

Page 16

Advertising
background image

16

Ammonia MR

SPECIFICATIONS

Range

0.00 to 10.00 mg/L

Resolution

0.01 mg/L

Accuracy

±0.05 mg/L ±5% of reading at 25 °C

Typical EMC

±0.01 mg/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 420 nm

Method

Adaptation of the

ASTM Manual of Water and Environmental Technology, D1426-92,

Nessler method. The reaction between ammonia and reagents causes a yellow tint in
the sample.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93715A-0

First Reagent

4 drops (6 drops for seawater)

HI 93715B-0

Second Reagent 4 drops (10 drops for seawater)

REAGENT SETS

HI 93715-01 Reagents for 100 tests
HI 93715-03 Reagents for 300 tests
For other accessories see page 46.

MEASUREMENT PROCEDURE

• Select the

Ammonia MR method using the procedure described in

the

Method Selection section (see page 12).

• Fill the cuvette with 10 mL of unreacted sample (up to the mark)

and replace the cap.

• Place the cuvette into the holder and close the lid.

• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for

measurement.

AMMONIA MEDIUM RANGE

10 mL