Hanna Instruments HI 736 User Manual

Page 2

Advertising
background image

• Turn the meter on by pressing the button.

After all the segments are displayed, “C.1”,
“Add” appears with “Press” blinking, the
meter is ready.

• Fill the cuvette with 10 mL of unreacted

sample and replace the cap. Place the cuvette
into the meter and close the meter’s cap.

• Press the button. When the display shows

“Add”, “C.2” with “Press” blinking the
meter is zeroed.

• Remove the cuvette, open it and add the

content of one packet of HI 736-25
reagent. Replace the cap and shake gently
for 2 minutes until the powder is completely
dissolved. Replace the cuvette into the
meter.

• Press and hold the button until the timer is

displayed on the LCD.

• The instrument directly displays the

concentration of phosphorus in ppb. The meter
automatically turns off after 2 minutes.

10 mL

• It is important that the sample does not contain any debris.
• Whenever the cuvette is placed into the measurement cell, it must by dry

outside, and completely free of fingerprints, oil or dirt. Wipe it throughly with
HI 731318 or a lint-free cloth prior to insertion.

• Shaking the cuvette can generate bubbles, causing higher readings. To obtain

accurate measurements, remove such bubbles by swirling or by gently tapping
the cuvette.

• Do not let the reacted sample stand for too long after reagent is added, or

accuracy will be lost.

• After the reading it is important to immediately discard the sample, otherwise

the glass might become permanently stained.

REAGENT SETS
HI 736-25

Reagents for 25 Phosphorus Ultra Low Range tests

OTHER ACCESSORIES
HI 740028

1.5V AAA batteries (4 pcs)

HI 731318

Tissue for wiping cuvettes (4 pcs)

HI 731321

Glass cuvettes (4 pcs)

HI 731353

Caps for cuvettes (4 pcs)

HI 93703-50

Cuvettes cleaning solution (230 mL).

Recommendations for Users
Before using these products, make sure that they are entirely suitable for your specific application and for
the environment in which they are used.
Operation of these instruments may cause unacceptable interferences to other electronic equipments, thus
requiring the operator to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the instrument’s EMC
performance.
To avoid damages or burns, do not put the instrument in microwave oven. For the safety of you and the
instrument do not use or store the instrument in hazardous environments.

For additional information, contact your

dealer or the nearest

Hanna Customer Service Center.

To find the Hanna Office in your area,

visit our web site

w w w . h a n n a i n s t . c o m

Hanna Instruments reserves the right to modify the design, construction and

appearance of its products without advance notice.

Battery management

To save the battery, the instrument shuts down after 3 minutes of non-use and 2 minutes
after read.
One fresh battery lasts for a minimum of 5000 measurements, depending on the light level.
When the battery capacity is under 10 % “bAt” appears on the LCD at start-up.
If the battery is empty and accurate measurements can’t be taken any more, the instrument
shows “bAd” then ”bAt” each for 1 second and turns off.
To restart the instrument, the battery must be replaced with a fresh one.
To replace the instrument’s battery, follow the steps:
• Turn the instrument off by holding the button until the meter shuts off.
• Turn the instrument upside down and remove the battery cover with a screwdriver.

• Remove the battery from its location and replace it with a fresh one.
• Insert the battery cover and replace the screw with a screwdriver.

IST736 05/10

Measurement procedure:

Accessories:

Tips for an accurate measurement

Advertising