pH/ORP/EC/Temperature in One Unit

HI 98204 Water Test

w w w . h a n n a i n s t . c o m

ISTWATERTSTR2 07/05

ON / OFF

SWITCH

RANGE SWITCH

LIQUID

CRYSTAL

DISPLAY

MAXIMUM

IMMERSION

LEVEL

BLACK

COMPARTMENT

FRONT

BATTERY

COVER

OFFSET & SLOPE

CALIBRATION

TRIMMERS

COVER

BACK

Water Test

ON / OFF

4 pH 7

RANGE

IMMERSION LEVEL

Water Test

ON / OFF

4 pH 7

RANGE

BATTERY

COVER

IMMERSION LEVEL

Recommendations for Users

Before using this product, make sure that it is entirely suitable for
the environment in which it is used. Operation of this instrument
in residential area could cause unacceptable interferences to
radio and TV equipments, requiring the operator to take all
necessary steps to correct interferences.

The glass bulb at the end of the pH electrode is sensitive to
electrostatic discharge. Avoid touching this glass bulb at all times.
For calibration it is recommended to use an antistatic
screwdriver.

Any variation introduced by the user to the supplied equipment
may degrade the instrument's EMC performance. To avoid
electrical shock, do not use this instrument when voltages at the
measurement surface exceed 24 Vac or 60 Vdc.

To avoid damages or burns, do not perform any measurement in
microwave ovens.

MAINTENANCE

• Separate the compartment from the tester.

• Remove deposits formed on the liquid junction

and pH glass bulb by soaking in HI 7061 cleaning
solution for 30 minutes.

• Soak the platinum wire in acetone for 30 minutes.

• Rinse glass bulb, platinum wire and liquid junc-

tion thoroughly with distilled water

• Re-assemble the compartment and the tester.

• Fill the compartment with tap water.

BATTERY REPLACEMENT

When the Water Test cannot be switched on or the
display fades, pull out the battery cover and replace
all three 1.5V batteries, while paying attention to their
polarity. Battery replacement must only take place in
a safe area and using the battery types specified in
this instruction manual.

ACCESSORIES

HI 7004M

pH 4.01 buffer solution, 230 mL

HI 7004L

pH 4.01 buffer solution, 500 mL

HI 7007M

pH 7.01 buffer solution, 230 mL

HI 7007L

pH 7.01 buffer solution, 500 mL

HI 7010M

pH 10.01 buffer solution, 230 mL

HI 7010L

pH 10.01 buffer solution, 500 mL

HI 7031M

1413 µS/cm conductivity solution, 230 mL

HI 7031L

1413 µS/cm Conductivity solution, 500 mL

HI 7061M

General cleaning solution, 230 mL

HI 7061L

General cleaning solution, 500 mL

HI 731326

Calibration screwdriver (20 pcs)

GENERAL DESCRIPTION

The Water Test is a "four-in-one" portable meter designed for quick
and reliable measurement of the four most important variables in
water analysis: pH, ORP, conductivity and temperature.

Water Test is engineered with great simplicity, that even non-techni-
cal personnel can use it: just fill the base with the sample to be
measured, press the RANGE key to select the mode and read the
measurement.

Water Test is perfect for use in the fields of water treatment analysis,
wastewater test, ecological studies, aquaculture and hydroponic ap-
plications.

SPECIFICATIONS

pH

ORP

Conductivity

Temperature

Range

0.0 to 14.0 pH

±1000 mV

0 to 1999 µS/cm

0.0 to 60.0°C

Resolution

0.1 pH

1 mV

1 µS/cm

0.1°C

Accuracy

±0.2 pH

±5 mV

±2% F.S.

±1°C

(@20°C/68°F)

Typical EMC

±0.4 pH

±2mV

±2% F.S.

±1°C

Deviation

Calibration

manual

factory

manual

factory

2 point

calibrated

1 point

calibrated

Battery Type/Life

3 x 1.5V / approx. 200 hours of continuous use

Environment

0 to 50°C (32 to 122°F); RH max 95%

Dimensions / Weight

190 x 85 x 85 mm (7.4 x 3.3 x 3.3") / 260 g (9.2 oz.)

Advertising