Hanna Instruments HI 4521 User Manual

Page 14

Advertising
background image

14

For each measurement mode (pH, mV, Rel mV, Ion, Conductivity, Resistivity, TDS or Salinity) the following
display configurations are available: Basic, Good Laboratory Practice (GLP), Graph and Log History.

Basic
Accessing this option, the measured value and its units are
displayed on the LCD, along with the temperature value, tem-
perature compensation mode, and minimal GLP data.
To choose the Basic displaying mode:

• Press

while in

Measure mode. The “Choose Display

Configuration” message will be displayed in the Reminder
messages area.

• Press

. The instrument will display the basic infor-

mation for the selected measurement mode.

GLP
Accessing this option, a detailed GLP data will be displayed on
the LCD for pH / ISE: Last Calibration Date and Time, Offset and
Slope values, Calibration Buffers, Electrode Condition and for
conductivity / salinity: Last Calibration Date and Time, Calibra-
tion Standards, Cell Constant, Probe Offset, Reference Temperature,
Compensation Coefficient, Temperature Compensation.
Note: If only a one-point pH calibration is performed or the current

calibration does not include at least two consecutive standard
buffers of pH 4.01, 7.01 (6.86) and 10.01 (9.18) buffers, the
Electrode Condition will be unknown.

To access the GLP displaying mode:

• Press

while in

Measure mode. The “Choose Display

Configuration” message will be displayed in the Reminder
messages area.

• Press

. The instrument will display the detailed GLP

data.

D I S P L A Y I N G M O D E S

D I S P L A Y I N G M O D E S

D I S P L A Y I N G M O D E S

D I S P L A Y I N G M O D E S

D I S P L A Y I N G M O D E S

103

PREPARATION PROCEDURE

Remove the protective cap of the pH electrode.
DO NOT BE ALARMED IF SALT DEPOSITS ARE PRESENT. This is normal with electrodes. They will disappear
when rinsed with water.
During transport, tiny bubbles of air may form inside the glass bulb affecting proper functioning of the electrode.
These bubbles can be removed by “shaking down” the electrode as you would do with a glass thermometer.
If the bulb and/or junction is dry, soak the electrode in

HI 70300 or HI 80300 Storage Solution for at least one hour.

For refillable electrodes:

For refillable electrodes:

For refillable electrodes:

For refillable electrodes:

For refillable electrodes:
If the filling solution (electrolyte) is more than 2½ cm (1”) below the fill hole, add

HI 7082 or HI 8082 3.5M

KCl Electrolyte Solution for double junction or

HI 7071 or HI 8071 3.5M KCl+AgCl Electrolyte Solution for

single junction electrodes.
For faster response, unscrew the fill hole screw during measurements.

E L E C T R O D E C O N D I T I O N I N G

E L E C T R O D E C O N D I T I O N I N G

E L E C T R O D E C O N D I T I O N I N G

E L E C T R O D E C O N D I T I O N I N G

E L E C T R O D E C O N D I T I O N I N G

& M A I N T E N A N C E

& M A I N T E N A N C E

& M A I N T E N A N C E

& M A I N T E N A N C E

& M A I N T E N A N C E

Advertising
This manual is related to the following products: