Hanna Instruments HI 2550 User Manual

Page 14

Advertising
background image

14

3 cm

(1¼")

FIVE

FIVE

FIVE

FIVE

FIVE-----POINT

POINT

POINT

POINT

POINT CALIBRATION

CALIBRATION

CALIBRATION

CALIBRATION

CALIBRATION

• Submerse the pH electrode and the temperature

probe approximately 3 cm (1¼”) into a buffer
solution and stir gently. The temperature probe
should be close to the pH electrode.

• Press CAL. The “CAL” and “

BUF

” tags will

appear and the “7.01” buffer will be displayed
on the secondary LCD.

• If necessary, press the ARROW keys to select a different buffer value.
• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected

buffer, the “READY” tag will be displayed and the
“CFM” tag will blink.

• Press CFM to confirm calibration.
• The calibrated value is then displayed on the primary

LCD and the second expected buffer value on the
secondary LCD.

• After the first calibration point is confirmed, submerse the pH

electrode and the temperature probe approximately 3 cm (1¼”)
into the second buffer solution and stir gently. The temperature
probe should be close to the pH electrode.

• If necessary, press the ARROW keys to select a different buffer value.
• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected

buffer, the “READY” tag will be displayed and the
“CFM” tag will blink.

• Press CFM to confirm calibration.
• The calibrated value is then displayed on the primary

LCD and the third expected buffer value on the
secondary LCD.

• After the second calibration point is confirmed, submerse the pH

electrode and the temperature probe approximately 3 cm (1¼”) into
the third buffer solution and stir gently. The temperature probe
should be close to the pH electrode.