Hanna Instruments HI 9147 User Manual

Hi 9147

Advertising
background image

Instruction Manual

HI 9147

Portable, Water-resistant

Dissolved Oxygen Meter

with

Galvanic Probe

and

Battery Recharging System

w w w . h a n n a i n s t . c o m

WARRANTY

HI 9147 meters are guaranteed for two years against

defects in workmanship and materials when used for their

intended purpose and maintained according to instructions.

The probes are guaranteed for one year.

This warranty is limited to repair or replacement free of charge.

Damages due to accident, misuse, tampering or lack of

prescribed maintenance are not covered.

If service is required, contact the dealer from whom you

purchased the instrument. If under warranty, report the model

number, date of purchase, serial number and the nature of the

failure.

If the repair is not covered by the warranty, you will be notified

of the charges incurred.

If the instrument is to be returned to Hanna Instruments, first

obtain a Returned Goods Authorization Number from the

Customer Service department and then send it with shipment

costs prepaid. When shipping any instrument, make sure it is

properly packaged for complete protection.

Dear Customer,
Thank you for choosing a Hanna product.
Please read carefully this instruction manual before using the

meter. If you need additional technical information, do not

hesitate to e-mail us at [email protected]

[email protected]

[email protected]

[email protected]

[email protected].

This instrument is in compliance with the CE directives.

PRELIMINARY EXAMINATION

Remove the instrument from the packing material and examine

it carefully to make sure that no damage has occurred during

shipment. If there is any damage, notify your Dealer.
The meter is supplied complete with

• D.O. probe, fixed:

HI 76409/4 with 4 m cable for HI 9147-04
HI 76409/10 with 10 m cable for HI 9147-10

• Spare membranes (HI 76409A/P) with O-rings, 5 pcs.
• HI 7042S electrolyte solution (30 ml bottle)
• HI 710016 shockproof rubber boot
• 9V Ni-MH rechargeable battery
• Screwdriver and instructions.

Note: Conserve all packing material until the instrument has

been observed to function correctly. Any defective item

must be returned in its original packing.

1) HI 76409/4 or HI 76409/10 galvanic D.O. probe (fixed)

2) Liquid Crystal Display
3) BL key, turn the backlight ON (as long as it is pressed)
4) °C key, display temperature measurement
5) % key, display O

2

measurement in % of saturation

6) % trimmer, for calibration adjustment
7) SAL trimmer, for setting salinity compensation value
8) SAL key, display setting of salinity factor (active only in

mg/L measurement mode)

9) mg/L key, display O

2

measurement in mg/L (ppm)

10) OFF key, turn the instrument OFF

Note: To turn the instrument ON, press any of the

measurement keys (°C, %, mg/L)

Range

0.0 to 50.0 mg/L (ppm) O

2

0 to 600 % O

2

-5.0 to 50.0°C

Resolution

0.1 mg/L or 1% (O

2

) / 0.1°C

Accuracy

±1% of reading (O

2

) / ±0.2°C

Typical EMC Deviation

±2% of reading (O

2

) / ±1°C

Calibration

Manual, in saturated air

Temperature Compensation

Automatic, from -5 to 50°C (23 to 122°F)

Salinity Compensation

0 to 51 g/L (resolution 1 g/L)

Probe (fixed)

HI 76409/4 with 4 m cable or
HI 76409/10 with 10 m cable

Battery Type

1 x 9V Ni-MH, rechargeable

Battery Charge Life

200 hours of use (BL off)

Battery Life

Approx. 5 years

Auto-off

After approx. 6 minutes

Environment

0 to 50°C (32 to 122°F);

max 95% RH non-condensing

Dimensions

164 x 76 x 45 mm

(6.5 x 3.0 x 1.8")

Weight

(with HI 76409/4 probe)

511 g (1.12 lb.)

ACCESSORIES

HI 76409/4* Galvanic D.O. probe with built-in temperature

sensor and 4 m cable

HI 76409/10*Galvanic D.O. probe with built-in temperature

sensor and 10 m cable

HI 76409-0 Protective sleeve for HI 76409 probes series
HI 76409A/P Membrane for HI 76409 probes series, 5 pcs.
HI 7042S

Electrolyte solution, 30 ml bottle

HI 710015 Shockproof rubber boot, blue color
HI 710016 Shockproof rubber boot, orange color
HI 731326 Calibration screwdriver (20 pcs)
HI 710040 Battery recharger

*

To be substituted by authorized technical personnel only

GENERAL DESCRIPTION

Oxycheck

Oxycheck

Oxycheck

Oxycheck

Oxycheck is a water-resistant Dissolved Oxygen meter with

manual calibration, automatic temperature compensation (ATC),

salinity compensation and direct probe.

The instrument is equipped with an inductive system for battery

recharge, with no external contact, to ensure a watertight seal.

Dissolved Oxygen is indicated in mg/L (ppm) or in % of saturation.

The temperature can be measured in the range from -5 to 50°C.

The Dissolved Oxygen readings are automatically compensated for

the temperature effects on the oxygen solubility and membrane

permeability. Moreover, the salinity compensation feature allows

determination of Dissolved Oxygen even in salty waters.

The meter is housed in a rugged water-resistant case for maximum

protection in the field as well as in the laboratory.

The D.O. probe is provided with a membrane covering the galvanic

sensors and a built-in thermistor for temperature measurement

and compensation. The thin permeable membrane isolates the

sensor elements from the testing solution, but allows oxygen to

enter. Oxygen that passes through the membrane causes a current

flow, from which the oxygen concentration is determined.

Two models are available:

• HI 9147-04 with HI 76409/4 probe (4 m cable)

• HI 9147-10 with HI 76409/10 probe (10 m cable)

IST9147R2 05/05

Hanna Instruments reserves the right to modify the design,

construction and appearance of its products without advance notice.

FUNCTIONAL DESCRIPTION

SPECIFICATIONS