Hanna Instruments HI 9141 User Manual

Page 12

Advertising
background image

23

22

PRINT

CAL

LOG

LOG

The printer will print a complete set of data
and the "LOG" symbol is displayed.

Each printout provides the following informa-
tion:
a - Current log number
b - Current sample number (in that log)
c - Date (DD-MM-YY)
d - Printing interval indicator in minutes
e - Time (HH-MM)
f - DO value in ppm
g - Temperature

value in °C

It is always possible to switch from the log-
ging with printing function to the logging func-
tion. Press the CAL key and the "LOG" sym-
bol will start blinking to indicate that the data
are now stored only into memory.

Notes:

It is recommended to use the external power
supply during logging with printing mode,
especially when many printouts are
required.

Before proceeding with logging with print-
ing, make sure there is enough paper for
your measurements. When the paper is
finished the meter will not advise the op-
erator and the printouts could be lost.
Anyway, data will continue to be stored
into the meter memory.

It is possible to insert a new paper roll
during logging session (see "Printer
Maintenance" section for details).

1 4 - 0 1 - 9 6

- - 0 3 - -

0 0 1 0 M

0 0 0 1

* 1 9 . 1 0

8 . 0 2 p p m

2 2 . 1 ° C

<>

a

b

c

d

e

f

g

Once in logging mode, the
interval cannot be changed.
First exit the logging mode
by pressing the LOG key,
and then set the new in-
terval.

If the PRINT key is
pressed while in logging
mode, a printout is pro-
duced without affecting the
running sample number.

TO STOP LOGGING
Press the LOG key, this will generate a log
exit status printout.

10-01-96

==07== 0005M

0016 *16.25
#1 20.0 °C

.

Log number

Total sample number

LOG EXIT
Date/Time

14-01-96

- -09- -

0001M

0016

*20.00

<>

Advertising
This manual is related to the following products: