Hanna Instruments HI 96727 User Manual

Page 2

Advertising
background image

14•Press GLP/ to edit the desired month

(01-12). If the key is kept pressed, the
month number is automatically increased.

15•When the correct month has been set,

press ZERO/CFM or READ/ to confirm.
Now the display will show the day blinking.

16•Press GLP/ to edit the desired day

(01-31). If the key is kept pressed, the
day number is automatically increased.

Note:

It is possible to change the editing from
day to year and to month by pressing
READ/ .

17•Press ZERO/CFM to save the calibration

date.

18•The instrument displays “Stor” for one

second and the calibration is saved.

19•The instrument will return automatically

to

measurement mode by displaying dashes

on the LCD.

OTHER ACCESSORIES

HI 96727-11

CAL CHECK™ Standard Cuvettes (1 set)

HI 721310

9V battery (10 pcs)

HI 731318

Tissue for wiping cuvettes (4 pcs)

HI 731331

Glass cuvettes (4 pcs)

HI 731335

Caps for cuvettes (4 pcs)

HI 93703-50

Cuvettes cleaning solution (230 mL).

REQUIRED ACCESSORIES
0.45

µ

m membrane for true color measurement.

Warranty

HI 96727

is warranted for two years against defects in workmanship and

materials when used for its intended purpose and maintained according
to the instructions.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact your dealer. If under warranty, report the
model number, date of purchase, serial number and the nature of the
failure. If the repair is not covered by the warranty, you will be notified
of the charges incurred.
If the instrument is to be returned to Hanna Instruments, first obtain a
Returned Goods Authorization Number from the Customer Service Department
and then send it with shipment costs prepaid. When shipping any
instrument, make sure it is properly packaged for complete protection.
To validate your warranty, fill out and return the enclosed warranty card
within 14 days from the date of purchase.

Recommendations for Users

Before using these products, make sure that they are entirely suitable for your specific application and for
the environment in which they are used.
Operation of these instruments may cause unacceptable interferences to other electronic equipments, this
requiring the operator to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the instrument’s EMC
performance.
To avoid damages or burns, do not put the instrument in microwave oven. For yours and the instrument
safety do not use or store the instrument in hazardous environments.

Hanna Instruments

reserves the right to modify the

design, construction and appearance of its products

without advance notice.

For additional information, contact your

dealer or the nearest

Hanna Customer Service Center.

To find the Hanna Office in your area,

visit our web site

w w w . h a n n a i n s t . c o m

IST96727 09/10

Accessories:

3

Warning:

do not validate or calibrate the instrument with standard

solutions other than the Hanna CAL CHECK™ Standards, otherwise
erroneous results will be obtained.
For accurate validation and calibration results, please perform tests at
room temperature (18 to 25°C; 64.5 to 77.0°F).

Use the Hanna CAL CHECK™ cuvettes (see
“Accessories”) to validate or calibrate instruments.

Validation and Calibration procedures

VALIDATION
1• Turn the meter on by pressing ON/OFF.
2• When the beeper sounds briefly and the

LCD displays dashes, the meter is ready.

3• Place the CAL CHECK™ Standard

HI 96727-11

Cuvette A into the holder

and ensure that the notch on the cap
is positioned securely into the groove.

4• Press ZERO/CFM and the lamp, cuvette

and detector icons will appear on the display,
depending on the measurement phase.

5• After a few seconds the display will show

“-0.0-”

. The meter is now zeroed and

ready for validation.

6• Remove the cuvette.
7• Place the CAL CHECK™ Standard

HI 96727-11

Cuvette B into the holder

and ensure that the notch on the cap is
positioned securely into the groove.

8• Press CAL CHECK key and the lamp,

cuvette and detector icons together with
“CAL CHECK”

will appear on the display,

depending on the measurement phase.

9• At the end of the measurement the display

will show the validation standard value.
The reading should be within specifications as
reported on the CAL CHECK™ Standard
Certificate. If the value is found out of
specifications, please check that the cuvettes
are free of fingerprints, oil or dirt and repeat
validation. If results are still found out of
specifications then recalibrate the instrument.

Validation

4-5

7

8

9

Calibration

3

4

5-6

8

9-10

11-13

or

14-15

or

16

17

FACTORY CALIBRATION RESTORE

It is possible to delete the calibration and restore
factory calibration.

1• Press GLP/ to enter

GLP mode.

2• Press READ/ to enter in the factory

calibration restore screen. The instrument
asks for confirmation of user calibration
delete.

LAST CALIBRATION DATE

1• Press GLP/ to enter

GLP mode. The

calibration month and day will appear on
the main display and the year on the
secondary display.

2• If no calibration was performed, the factory

calibration message, “F.CAL” will appear
on the main display and the instrument
returns to

measurement mode after three

seconds.

GLP

Last

Calibration

Date

In

GLP mode, the last calibration date can be verified and the factory

calibration can be restored.

Factory

Calibration

Restore

2

3-4

Battery management

To save the battery, the instrument shuts down after 10 minutes of non-use
in

measurement mode and after 1 hour of non-use in calibration mode.

If a valid measurement was displayed before
auto-shut off, the value is displayed when the
instrument is switched on. The blinking “ZERO”
means that a new zero has to be performed.
One fresh battery lasts for arround 750 measurements, depending on the
light level.
The remaining battery capacity is evaluated at the instrument startup
and after each measurement.
The instrument displays a battery indicator with three levels as follows:
• 3 lines for 100 % capacity
• 2 lines for 66 % capacity
• 1 line for 33 % capacity
• Battery icon blinking if the capacity is under 10 %.
If the battery is empty and accurate measurements can’t be taken any
more, the instrument shows “dEAd bAtt” and turns off.
To restart the instrument, the battery must be replaced with a fresh one.
To replace the instrument’s battery, follow the steps:
• Turn the instrument off by pressing ON/OFF.
• Turn the instrument upside down and remove the battery cover by

turning it counterclockwise.

• Extract the battery from its location and replace it with a fresh one.
• Insert back the battery cover and turn it clockwise to close.

CALIBRATION
Note:

It is possible to interrupt the calibration
procedure at any time by pressing CAL
CHECK

or ON/OFF keys.

1• Turn the meter on by pressing ON/OFF.

2• When the beeper sounds briefly and the

LCD displays dashes, the meter is ready.

3• Press and hold CAL CHECK for three seconds

to enter

calibration mode. The display will

show “CAL” during calibration procedure. The
blinking “ZERO” asks for instrument zeroing.

4• Place the CAL CHECK™ Standard

HI 96727-11

Cuvette A into the cuvette

holder and ensure that the notch on the cap
is positioned securely into the groove.

5• Press ZERO/CFM and the lamp, cuvette

and detector icons will appear on the display,
depending on the measurement phase.

6• After a few seconds the display will show

“-0.0-”

. The meter is now zeroed and

ready for calibration. The blinking “READ”
asks for reading calibration standard.

7• Remove the cuvette.

8• Place the CAL CHECK™ Standard

HI 96727-11

Cuvette B into the holder

and ensure that the notch on the cap is
positioned securely into the groove.

9• Press READ/ and the lamp, cuvette and

detector icons will appear on the display,
depending on the measurement phase.

10•The instrument will show for three seconds

the CAL CHECK™ standard value.

Note:

If the display shows “STD HIGH”, the
standard value was too high. If the
display shows “STD LOW”, the standard
value was too low. Verify that both CAL
CHECK™

Standard HI 96727-11 Cu-

vettes, A and B are free from fingerprints
or dirt and that they are inserted cor-
rectly.

11•Then the date of last calibration (e.g.:

“01.08.2009”

) appears on the display, or

“01.01.2009”

if the factory calibration

was selected before. In both cases the year
number is blinking, ready for date input.

12•Press GLP/ to edit the desired year

(2008-2099). If the key is kept pressed,
the year number is automatically increased.

13•When the correct year has been set, press

ZERO/CFM

or READ/ to confirm. Now

the display will show the month blinking.

14•Press READ/ and the lamp, cuvette

and detector icons will appear on the
display, depending on the measurement
phase.

15•The meter directly displays concentration

in color units of the true color on the LCD.

INTERFERENCES:
There are no known interferences for this method.

3• Press ZERO/CFM to restore the factory

calibration or press GLP/ again to
abort factory calibration restore.

4• The instrument briefly indicates “donE”

upon restoration of factory calibration prior
to returning to

measurement mode.

2

1

18

15

14